Table 5. Troubleshooting (cont.) 

Symptom 

Possible cause 

Action required 

Volume of residual 

liquid in a well 

exceeds the 

admissible value (2 

μl) 

Depth of the manifold 

immersion is not 

determined. 

Perform well depth measurement in the automatic 

mode. 

Obstruction of the manifold 

aspiration channel or a 

needle. 

Check if the aspiration channel and needles are not 

obstructed. If yes, do their cleaning (see 

“Obstruction of dispensing channel or a needle in 

the manifold” in this Table). 

If it does not help, turn off the device and have it 

checked by a specialist. 

On starting the 

operation, the device 

does not fetch liquid 

from the active 

bottle. 

Insufficient liquid quantity 

in the pipe. 

To solve this problem, complete the following 

procedure: 

- Stop the program. 

- Prepare a syringe with 5 ml of distilled water. 

- Unscrew the hose from the active bottle. 

- Press Prime Rinse key. 

- Using + and - keys, select number of the active 

bottle and press the Start key. 

- At the moment when the pump begins operation, 

gradually fill the hose with distilled water from the 

syringe. 

- Tighten the hose on the bottle. 

If the pump does not restore performance, contact 

the service engineer  

10.8. 

Storage and transportation

10.8.1. Remove the manifold, tubes and platform for plates. 
10.8.2. Put  on  packaging  blocks  onto  sides  of  the unit  and place  it  into  the  original carton 

box. 

10.8.3. Place the manifold into protecting case; place syringe, manifold cleaning set and the 

platform into PE bag. Place the bag and the case with the manifold in the corners of 
the box between the packaging blocks and the side panels of the carton box; place 
the 2-litre bottle and the half-litre bottle in free space between the packaging blocks 
and the side panels of the carton box. 

10.8.4. Insert the top section into the carton box by bending the section downwards along 

the perimeter and inserting the bent parts between the walls of the original carton box 
and packaging blocks, so that the device and the remaining content of the box are 
fully covered. 

10.8.5. Put the soft foam sheet (grey) with the smooth side down on the resulting horizontal 

carton surface.  

10.8.6. Put the remaining three bottles and other accessories on the foam sheet. 
10.8.7. If 4-channel weight logger is included in the kit, insert pads under scale cups, then 

place the logger into top section of packaging. 

10.8.8. Cover everything with another sheet of soft foam material with the smooth side up. 
10.8.9. Close the box and seal with packing tape.  

Содержание IW-8

Страница 1: ......

Страница 2: ...2...

Страница 3: ...Program error messages 16 8 Specifications 16 9 Ordering information 17 10 Maintenance and troubleshooting 18 11 Warranty 21 12 Glossary 22 13 EU Declaration of conformity 23 1 About this edition of...

Страница 4: ...rd the results obtained and repeat the washing procedure Accuracy of the obtained testing results depends on the quality of the microplate rinsing procedure To assure correctness of the rinsing proced...

Страница 5: ...erflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory standards 2 4 AFTER...

Страница 6: ...W 8 the unit provides automatic control of reagent total volume in all bottles that is necessary for washing and waste collecting and estimates the volume of reagents necessary for washing Microbiolog...

Страница 7: ...bes outside inside diam length 6 3 600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hy...

Страница 8: ...in a dry dust free place If the unit will be subjected to low temperatures dry the tubes as described in 6 Operation closedown Use the original packaging for transportation whether transported by air...

Страница 9: ...4 4 6 Install the manifold in the slots of the holder arm fig 2 3 Magnets in the manifold and the holder arm connect and align the manifold 4 4 7 Connect the manifold inlet connector supplying the li...

Страница 10: ...ce bottles near the unit 4 4 19 If 4 channel weight logger is included in the kit remove pads from underside of scale cups Place bottles on the 4 channel weight logger according to the numbering Conne...

Страница 11: ..._30 5 1x400_30 Soak 5 min Dispense 200 Dispense NO YES YES YES YES YES YES YES YES Aspirate YES YES YES YES YES YES YES YES YES Shake NO NO NO NO NO NO NO NO NO Dispense rate 02 02 02 02 02 02 02 02 0...

Страница 12: ...ttle to use for the main washing Second chan cycles 01 15 Number of washes with second solution 2 channel 1 3 Number of bottle to use for the second washing 5 4 Install the plate on the platform Atten...

Страница 13: ...y bottle number using and keys and press the Run Stop key 5 16 During operation display shows program number currently used bottle number and the number of cycles left fig 6 If the current action has...

Страница 14: ...section 6 Operation closedown of this manual Attention Always perform the rinsing cycle using distilled water after finishing the operation at the end of working day This will help to keep the unit in...

Страница 15: ...l figure 3 6 2 4 Press the Run Stop key fig 4 7 and the system performs a rinsing cycle Repeat the procedure twice if necessary 6 2 5 Repeat 6 2 2 6 2 4 for the remaining tubes 6 3 After finishing the...

Страница 16: ...ration in cold rooms and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Страница 17: ...IW 8 Intelispeed Washer V 4AW BS 060106 AAI 9 2 To inquire about or order the optional accessories or the replacement parts contact Biosan or your local Biosan representative 9 2 1 Optional accessori...

Страница 18: ...m the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring...

Страница 19: ...sed by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean ne...

Страница 20: ...service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original carton bo...

Страница 21: ...ddress To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For IW 8 the Premium class model one year of extended warranty is available free of cha...

Страница 22: ...ich provides solution aspiration consequently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottle fo...

Страница 23: ...l equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General...

Страница 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 02 November of 2020...

Отзывы: