14
Note.
If operation requires washing from one channel (parameter “On two
channels” set to NO), washing parameters for the second channel are
not displayed. Respectively, if two
channels are set (parameter “On
two channels” set to YES), washing parameters for the second
channel are displayed and can be adjusted as needed.
Figure 7. Display during operation with
connected 4-CHW Logger
Figure 8. Choosing the bottle number for rinse
on the second channel
5.19. When the cycle is completed, unit produces an informative sound signal. Proceed to
section
6. Operation closedown
of this manual.
Attention!
Always perform the rinsing cycle using distilled water after finishing
the operation at the end of working day. This will help to keep the unit
in working order and prevent head channel clogging.
5.20.
Plate setup
. When installing a plate for the first time or when a plate of different type
is used, setup the manifold needles lowering depth.
5.20.1. Press the
Enter/Calibr.
key in the start mode, the following message appears on the
display (fig.9).
5.20.2. To calibrate, press
Run/Stop
key. The manifold measures the depth of the plate well
by touching the plate surface first and then the bottom of the well. The unit saves the
difference of the measured values.
Figure 9. Plate setup