5.  Operation 

5.1. 

Before starting plate washing: 

Fill the rinsing liquid bottle (#3) with distilled water or appropriate cleansing solution. 

Fill the necessary bottles (#1 and/or #2) with washing solutions. 

5.2. 

Initial parameters of all programs are given in Table 1. All values are adjustable. User 
programs can be inputted as 00-100. User programs can be created by changing the 
template of the original program. 

Table 1. Initial parameters of all user programs. 

Program # 

00 

01 

02 

03 

04 

05 

06 

07 

08-100 

Parameter 

A

s

p

ira

te

 

1

x

4

0

0

_

3

0

 

2

x

4

0

0

_

3

0

 

3

x

4

0

0

_

3

0

 

5

x

4

0

0

_

3

0

 

5

+1

x

4

0

0

_

3

0

 

S

o

a

k

 5

 m

in.

 

D

is

p

e

n

s

e

 200

 

 

Dispense 

NO 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

Aspirate 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

Shake 

NO 

NO 

NO 

NO 

NO 

NO 

NO 

NO 

NO 

Dispense rate 

02 

02 

02 

02 

02 

02 

02 

02 

03 

Aspirate rate 

03 

03 

03 

03 

03 

03 

03 

03 

03 

Soak time, s 

20 

30 

30 

30 

30 

30 

300 

00 

00 

Shake time, s 

30 

30 

Double aspiration 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

YES 

Dispensed volume, μl 

400 

400 

400 

400 

400 

300 

200 

300 

Aspiration time, ms 

1200 

600 

600 

600 

600 

600 

600 

1000 

Final aspirate, ms 

1200  1200  1200  1200  1200  1200 

YES 

First aspirate 

YES 

YES 

YES 

YES 

YES 

NO 

YES 

Wash by rows 

YES 

YES 

YES 

YES 

YES 

NO 

YES 

On two channels 

NO 

NO 

NO 

NO 

YES 

NO 

NO 

Num.of 1 wash cycles 

01 

02 

03 

05 

05 

01 

03 

1 channel 

01 

01 

01 

01 

01 

01 

01 

01 

Num.of 2 wash cycles 

01 

2 channel 

02 

 

 

 

Содержание IW-8

Страница 1: ......

Страница 2: ...2...

Страница 3: ...Program error messages 16 8 Specifications 16 9 Ordering information 17 10 Maintenance and troubleshooting 18 11 Warranty 21 12 Glossary 22 13 EU Declaration of conformity 23 1 About this edition of...

Страница 4: ...rd the results obtained and repeat the washing procedure Accuracy of the obtained testing results depends on the quality of the microplate rinsing procedure To assure correctness of the rinsing proced...

Страница 5: ...erflow the maximum level during operation there is a mark on the side surface of the bottle Take necessary precautions utilizing waste liquid in accordance with general laboratory standards 2 4 AFTER...

Страница 6: ...W 8 the unit provides automatic control of reagent total volume in all bottles that is necessary for washing and waste collecting and estimates the volume of reagents necessary for washing Microbiolog...

Страница 7: ...bes outside inside diam length 6 3 600 mm 5 pieces 7 Tube for manifold outside inside diam length 3 2 1 6 400 mm 1 piece 8 Tube for manifold outside inside diam length 5 3 440 mm 1 piece 9 Tube for hy...

Страница 8: ...in a dry dust free place If the unit will be subjected to low temperatures dry the tubes as described in 6 Operation closedown Use the original packaging for transportation whether transported by air...

Страница 9: ...4 4 6 Install the manifold in the slots of the holder arm fig 2 3 Magnets in the manifold and the holder arm connect and align the manifold 4 4 7 Connect the manifold inlet connector supplying the li...

Страница 10: ...ce bottles near the unit 4 4 19 If 4 channel weight logger is included in the kit remove pads from underside of scale cups Place bottles on the 4 channel weight logger according to the numbering Conne...

Страница 11: ..._30 5 1x400_30 Soak 5 min Dispense 200 Dispense NO YES YES YES YES YES YES YES YES Aspirate YES YES YES YES YES YES YES YES YES Shake NO NO NO NO NO NO NO NO NO Dispense rate 02 02 02 02 02 02 02 02 0...

Страница 12: ...ttle to use for the main washing Second chan cycles 01 15 Number of washes with second solution 2 channel 1 3 Number of bottle to use for the second washing 5 4 Install the plate on the platform Atten...

Страница 13: ...y bottle number using and keys and press the Run Stop key 5 16 During operation display shows program number currently used bottle number and the number of cycles left fig 6 If the current action has...

Страница 14: ...section 6 Operation closedown of this manual Attention Always perform the rinsing cycle using distilled water after finishing the operation at the end of working day This will help to keep the unit in...

Страница 15: ...l figure 3 6 2 4 Press the Run Stop key fig 4 7 and the system performs a rinsing cycle Repeat the procedure twice if necessary 6 2 5 Repeat 6 2 2 6 2 4 for the remaining tubes 6 3 After finishing the...

Страница 16: ...ration in cold rooms and closed laboratory rooms at ambient temperature from 4 C to 40 C in a non condensing atmosphere and maximum relative humidity 80 for temperatures up to 31 C decreasing linearly...

Страница 17: ...IW 8 Intelispeed Washer V 4AW BS 060106 AAI 9 2 To inquire about or order the optional accessories or the replacement parts contact Biosan or your local Biosan representative 9 2 1 Optional accessori...

Страница 18: ...m the rinsing cycle twice using distilled water after finishing the operation see section 5 10 4 2 Monthly With 75 ethanol solution clean mounting surface of the plate platform and plate holder spring...

Страница 19: ...sed by needles they may contain dangerous infections Take off the manifold from the device put it down with needles upward Clean the aspiration channel needles with the wiping rod 0 80 mm and clean ne...

Страница 20: ...service engineer 10 8 Storage and transportation 10 8 1 Remove the manifold tubes and platform for plates 10 8 2 Put on packaging blocks onto sides of the unit and place it into the original carton bo...

Страница 21: ...ddress To obtain the claim form visit Technical support page on our website at link below 11 5 Extended warranty For IW 8 the Premium class model one year of extended warranty is available free of cha...

Страница 22: ...ich provides solution aspiration consequently from two opposite points of a well Strip washing mode plate washing by rows Dispensing filling wells with a set solution volume Waste bottle the bottle fo...

Страница 23: ...l equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements General...

Страница 24: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 4 02 November of 2020...

Отзывы: