- Before using the
se, please read all the instructions and
practice the tests including quality control test. ( Refer to page 46.)
- Please do the quality control test regularly to make sure the test results are accurate .
- The
the
under any circumstances. The use of
other brands of strips may give inaccurate results.
- The
for
in vitro
diagnostic use only. The
testing result is calibrated to be plasma equivalent for test with fresh capillary whole blood samples
by finger stick.
- The
for self-testing. It should not be
used to diagnose diabetes mellitus.
-
idated for use on neonates.
Therefore, it's not intended for use of neonates.
System to test your blood gluco
Blood Glucose Meter can only be used with
Blood Glucose Test
Strips. The use of any other brand Strips should not be used
Blood Glucose Monitoring System is intended
Blood Glucose Monitoring System is intended
Blood Glucose Monitoring System has not been val
Precaution
5
4
Precaution
- Blood Glucose Meter hasn't been used for arterial blood testing yet, so it is still not
suitable to be used for arterial blood testing so far.
- Do the test at least 30 minutes after while entering another site with different ambient temperature.
- Pay attention to the environment protection when disposal of batteries.
- Using appliance in tropical and / or temperate climate.
- Not to be exposed dripping or splashing.
- The minimum blood sample size of Glucose Monitoring System is 1.4µl :( )
1.0 1.4 2.0 3.0
µl
µl
µl
µl
4 µl
.0
We suggest you to take 1.4~2.5µl to do the test on glucose monitoring system. Blood sample size
above 4.0µl might contaminate the Smart Code Key.
Blood sample size below 1.4µl might cause inaccurate test result or might not start on meter
measurement. In this case, repeat the test with a new test strip.
Sample Size Example