4. The minimum blood sample size of Glucose Monitoring System is 1.4µl :( )
PRECAUTION
- Don’t take too much blood sample, such as 4.0µl ( please see how much it is in the
example picture ) or more, to do the test. Too much blood might contaminate the Smart
Code Key.
- Please clean it up right away if your Smart Code Key gets blood stain or moisture.
( Refer to page 55 for cleaning Smart Code Key )
- Blood sample size below 1.4µl might cause inaccurate test result or might not start on
meter measurement. In this case, repeat the test with a new test strip.
36
1. Touch the blood drop
to the sample entry of
the strip. The blood
will be drawn into the
strip automatically.
2. Hold your finger on the sample entry until you hear a " beep " and
the view window is totally filled with blood. If the view window is not
totally filled with blood and/or the test does not start, discard the
test strip and repeat the test with a new test strip. When you repeat
the test, you may want to obtain the blood sample from a different
finger if the first attempt did not produce enough blood sample.
PRECAUTION
- When you do a test, please pay attention not to squeeze too much blood to pollute Smart
Code Key.
37
Performing a Test in Order
Performing a Test in Order
We suggest you to take 1.4~2.5µl to do the test on glucose monitoring system. Blood sample size
above 4.0µl might contaminate the Smart Code Key.
1.0 1.4 2.0 3.0
µl
µl
µl
µl
4 µl
.0
Sample Size Example
Applying blood sample to test strip