8. When you hear a sound of " beep " , leave the meter on the table while waiting
the test result. The screen will show the countdown mode.
9. Tightly recap the cap of control solution.
10. The control result appears after the measurement is completed. Compare your
control test result to the range printed on strip vial label.
Your result should fall within the control solution
range printed on the label of strip vial. If result is
not within the control range, please refer to
Understanding Control Test Results
on page 54.
Please follow the local regulation and discard the
used strip properly.
Performing a Quality Control Test
Performing a Quality Control Test
PRECAUTION
- Your quality control test result by control solution test will not be calculated for average
reading but still can be recalled. The control solution test result will be shown with " "
symbol on the screen.
- You have to do check key test before control solution test. If you only use the control
solution to do the quality control test, your test result will be memorized and calculated for
average same as real blood test.
- Please don't touch the control solution by approaching the vial to
the sample entry on strip ( see drawing ). The reagent might be
sucked into the vial of control solution which alteration or
degeneration might happen. If you approach the control solution
vial to the sample entry on strip for 3~5 times, you will find that the
measurements might lower down for 10-20%.
- Keep your Smart Code Key dry and clean. Please clean it up right
away if your Smart Code Key gets stain or moisture.
( Refer to page 55 for cleaning )
Example
CLOSE
Control Solution Range
90-130 mg/dL
5.0-7.2mmol/L