Error Messages and Trouble Shooting
59
58
About early applying blood sample or control solution
About Strip Error - Er1
About Meter Error - Er2
About Signal Error - Er3
The blood sample or control solution are applied before the meter is ready. Repeat
the test with a new test strip. Add sample after the " " symbol appears.
Test strip inserted has been used or damaged. Please use a new test strip from vial.
Meter has malfunctioned. Do the quality control test or reinstall the batteries to see if
the meter works properly.
If error screen still appears, contact Bionime Customer Service.
When you do the test, the meter find the signal is unreasonable. Please do a new
test again.
Limitations
- Store the strips in the original capped vial at temperatures between 4
℃
to 30
℃
( 39
℉
to 86
℉
) and
relative humidity below 90%. Do not freeze.
- Blood Glucose Test strips are designed for using with capillary whole blood and venous
whole blood samples. Do not use serum or plasma samples.
- Inaccurate test results may be obtained at high altitude more than about 10000 feet ( 3048 meters )
above sea level.
- Hematocrit below 30% may cause higher results, and hematocrit above 55% may cause lower
results.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Blood Glucose Monitoring System has not been validated for use on neonates.
- The glucose test may be interfered under abnormal concentration of
Uric acid > 0.54 mmol/L ( > 9.0 mg/dL )
L - Dopa > 0.076 mmol/L ( > 1.5 mg/dL )
Methyldopa > 0.071 mmol/L ( > 1.5 mg/dL )
Cholesterol > 6.5 mmol/L ( > 250 mg/dL )