Error Messages and Trouble Shooting
About Temperature Error
1. In order to get accurate testing, please do testing between 10~40
℃
( 50~104
℉
).
2. When the ambient temperature is 0~9
℃
( 32~48
℉
) or 41~50
℃
(106~122
℉
) ,
the " " warning symbol will blinking, you still can do the test but the result
obtained is only for reference because the test result under these range of
temperature might not be correct. Repeat the test at an area with temperature
between operating range. ( 10~40
℃
or 50~104
℉
)
3. When you move from the area with temperature outside the operating range of
test strip to another area with temperature under operating range of test strip,
please wait for 30 minutes before you do the test.
4. When the temperature is below 0
℃
( 32
℉
) or over 50
℃
( 122
℉
), Meter can not
do the test and the " " symbol will blink in this condition.
Please move the meter to environment with temperature between 10~40
℃
( 50~104
℉
) and redo the test after 30 minutes.
Error Messages and Trouble Shooting
About Battery Error
1. The " " symbol is blinking when the battery power is low. Please change
batteries as soon as you can. You can still do the test.
2. The " " and " " symbol are blinking when the battery is too low. Meter
can not do the strip test. Please change the batteries immediately.
3. After changing the batteries you should do the quality control test.
About Check Key Error
When you see " " and " " symbols showing on screen, please do the
check key test again. If this error message appears again, it would be Meter or
Check Key defect.
Please return your Meter and Check Key to the store where you bought.
About Smart Code Key Error
The Smart Code Key is faulty or not assembled well to the meter. Reinsert the Smart
Code Key. If the error screen appears again, contact Bionime Customer Service.