C7SM-1.01
25/70
Service Manual of Cardio 2000
2-5. Trouble Shooting Guide
1) Power is off
The power is off
Does the
power input
cable break
down?
Change the power
input cable
Is the fuse
blown?
Change the fuse
Again blown
fuse
Does the
power switch
cable short-
circuit?
Change the power
switch cable
Does the main
power
connector
break down?
Change the main
power connector
Normal
operation
Use
No
Yes
Yes
Yes
Yes
No
Yes
B
A
B
Does the main power
module produce the
power properly?
Repair or change
the main power
module
No
Yes
Ask the
manufacturer
about it, then take
action
No
Yes
A
A
A
A
A
No
No
No