C7SM-1.01
28/70
Service Manual of Cardio 2000
4) EKG wave does not come out or it's very small
EKG wave does
not come out or
small
Is the patient
cable
connecter
correctly?
Reconnect the
patient cable
carefully
Is the patient
cable short-
circuited?
Is the
electrode
connected to
the patient
cable
correctly?
Normal
operation
Use
No
No
Yes
Yes
B
A
B
Is the
electrode
connected to
the patient
cable
correctly?
A
Change the
patient cable
A
Cleanse the
electrode with
alcohol or water,
then it use
Reconnected the
electrode to the
patient cable
properly
Ask the
manufacturer
about it, then take
action
No
A
A
Yes
No
Yes
Control the
volume by setting
a gain on the
main board
Is the EKG
signal adjusted
properly?
No
Yes
A
Содержание CardioCare EKG2000
Страница 1: ...EKG2000 SERVICE MANUAL Version 1 0 www ebionet com ...
Страница 47: ...C7SM 1 01 47 70 Service Manual of Cardio 2000 2 Cable LCD Assembly 40Pin FFC Cable Battery Cable Assembly ...
Страница 48: ...C7SM 1 01 48 70 Service Manual of Cardio 2000 QN TPH Cable Motor Assembly Power Switch Cable Assembly ...
Страница 51: ...C7SM 1 01 51 70 Service Manual of Cardio 2000 Bottom Cover Assembly Top Cover Assembly ...
Страница 61: ...C7SM 1 01 61 70 Service Manual of Cardio 2000 Bottom Cover Ass y Printer Module Ass y ...
Страница 62: ...C7SM 1 01 62 70 Service Manual of Cardio 2000 EKG2000 Ass y Battery Ass y ...
Страница 65: ...C7SM 1 01 65 70 Service Manual of Cardio 2000 8 2 CDUI Board Top Bottom ...
Страница 66: ...C7SM 1 01 66 70 Service Manual of Cardio 2000 8 3 CPU Board Top Bottom ...