Service Manual of Cardio 2000
2-4. Filter
Base line Filter
Base line drift is caused by a patient‟s breathing. ECG signal is overlapped on a large parabola. If
you want to set Base line drift, select „ON‟. Otherwise, select OFF. The set value appears on the
bottom part of the printed form indicating 0.1HZ- when set to ON, and 0Hz- when set to OFF.
<BASE FILTER OFF>
<BASE FILTER ON>
AC Filter
In addition to ECG, other signals such as power noise, base line drift by respiration and
electromyogram may be recorded in the output signals.
There are three types of the noise filter; 60Hz, 50Hz and OFF. When set to OFF, noise is not
eliminated. Set to 50Hz, it indicates to filter 50Hz power noise. Set to 60Hz, it means filtering 60Hz
power noise.
When the system is powered by battery, power noise is not generated, so you may have clear
ECG signal if set to OFF. In Europe where the power noise is 50Hz, the system is set to 50Hz. In other
nations including Korea and the U.S., it is set to 60Hz.
The status of the set filter can be checked on the bottom part of the printout, indicating AC 60Hz,
AC 50Hz or AC OFF.
<AC FILTER OFF>
Содержание CardioCare EKG2000
Страница 1: ...EKG2000 SERVICE MANUAL Version 1 0 www ebionet com ...
Страница 47: ...C7SM 1 01 47 70 Service Manual of Cardio 2000 2 Cable LCD Assembly 40Pin FFC Cable Battery Cable Assembly ...
Страница 48: ...C7SM 1 01 48 70 Service Manual of Cardio 2000 QN TPH Cable Motor Assembly Power Switch Cable Assembly ...
Страница 51: ...C7SM 1 01 51 70 Service Manual of Cardio 2000 Bottom Cover Assembly Top Cover Assembly ...
Страница 61: ...C7SM 1 01 61 70 Service Manual of Cardio 2000 Bottom Cover Ass y Printer Module Ass y ...
Страница 62: ...C7SM 1 01 62 70 Service Manual of Cardio 2000 EKG2000 Ass y Battery Ass y ...
Страница 65: ...C7SM 1 01 65 70 Service Manual of Cardio 2000 8 2 CDUI Board Top Bottom ...
Страница 66: ...C7SM 1 01 66 70 Service Manual of Cardio 2000 8 3 CPU Board Top Bottom ...