C7SM-1.01
26/70
Service Manual of Cardio 2000
2) LCD is blank although the power is on
LCD is blank
although the
power is on
Broken 20 pin
shield cable?
Change the 20
pin shield cable
Does the LCD cable
have a poor condition
of connection or short-
circuit?
Check the
condition of the
connection and
then change if
it's not good
Is the LCD
module
unsatisfactory?
Change the
LCD Module
Normal
operation
Use
No
Yes
Yes
Yes
No
Yes
B
A
B
Ask the
manufacturer
about it, then take
action
A
A
A
No
No