Rev. 1.1 5.ECG 93
BM3VET Touch User’s Manual
5.2 ECG Data Window
Note
ECG Wave Display is always on when the cable is connected.
The heart rate is calculated by a moving average. The monitor detects 8 consecutive beats,
averages the R-R intervals of the latest 8 beats and uses this average to calculate the current
heart rate. When a new beat is detected, the heart rate is recalculated using the latest 8beats. The
heart rate display is updated every 3 seconds.
Heart rate meter updates a new heart rate for a step increase or decrease in 10 seconds maximum.
When ventricular tachycardia is detected, the alarm set in 5 seconds maximum.
Check that the delay time of the output signal (alarm trigger 80ms maximum) is within the range of
the connected equipment.
80
50
100
P
BPM
PVC
---
Heart Rate Alarm
Limit:
Decides the
QRS limits, and gives
an alarm if a value is
over the limits.
QRS:
Detects QRS, and
flashes when QRS is
detected.
Heart Rate:
Displays heart rate per minute.
Pace Detector
Indicators:
Detects and displays
the Animal
’s pace
maker and flashes.
Heart Rate Alarm
Limit:
Decides the
QRS limits, and gives
an alarm if a value is
over the limits.
PVC Count:
Detects
and displays the
patient
’s PVC rhythm
Counting.
Содержание BM3VET Touch
Страница 1: ...BM3VET Touch User s Manual Veterinary Patient Monitor Rev 1 1 20160401...
Страница 29: ...Rev 1 1 1 BASIC 29 BM3VET Touch User s Manual Features of Main Body FRONT BACK...
Страница 78: ...Rev 1 1 4 TREND 78 BM3VET Touch User s Manual 4 TREND 4 1 TREND GRAPHIC TREND TABLE TREND TREND WINDOW SETUP...
Страница 87: ...Rev 1 1 5 ECG 87 BM3VET Touch User s Manual IEC 3LEAD CABLE AHA 3LEAD CABLE IEC 5LEAD CABLE AHA 5LEAD CABLE...
Страница 88: ...Rev 1 1 5 ECG 88 BM3VET Touch User s Manual IEC 3LEAD AHA 3LEAD IEC 5LEAD AHA 5LEAD...
Страница 91: ...Rev 1 1 5 ECG 91 BM3VET Touch User s Manual 5 Position of 5 Lead RL V LL LA RA...
Страница 92: ...Rev 1 1 5 ECG 92 BM3VET Touch User s Manual Position of 3 Lead Wire Electrodes LL LA RA...