14. TROUBLE SHOOTING
180
BM3VET Touch User’s Manual
14.4 NIBP malfunction
Are leaks from the hose Yes
connector of cuff
No
Repair the NIBP B/D
14.5 Abnormality in NIBP measurements
The Animal moved Yes
while measuring
No
The measurements Yes
appear lower
No
There is a delay Yes
in measuring
No
Repair the NIBP B/D
Replace the hose of cuff
Loosen the cuff during diagnosis
Tighten the cuff during diagnosis
Make sure the Animal remains
still during diagnosis