Rev. 1.1 8. NIBP 130
BM3VET Touch User’s Manual
CUFF SIZE
The user can select a CUFF between ADULT and NEONATAL.
NIBP
X
LARGE
(5<)
MEDI.
(3-4)
SMALL
(1-2)
CUFF SIZE:
LARGE
ALARM
INFLATION:
170mmHg
INTERVAL:
OFF
UNIT
SELECT:
mmHg
NIBP
STAT:
OFF
NIBP
VITAL SIGN
INFLATION
It is a function for set the maximum initial inflation pressure value.The range of initial inflation
pressure value of BM3VET TOUCH is as follows.
LARGE : 120
– 250 mmHg / MEDI. : 120 – 250mmHg / SMALL : 60 – 140mmHg
Default Value: LARGE : 170 mmHg / MEDI. : 140mmHg / SMALL : 120mmHg
NIBP
X
200
190
180
170
140
160
150
130
120
230
220
210
250
240
CUFF SIZE:
LARGE
ALARM
INFLATION:
170mmHg
INTERVAL:
OFF
UNIT
SELECT:
mmHg
NIBP
STAT:
OFF
NIBP
VITAL SIGN