Rev. 1.1 5.ECG 107
BM3VET Touch User’s Manual
Trouble shooting
Problem :
Inaccurate heart rate and/or false a systole.
Solution :
Check ECG signal from Animal:
1. Check/adjust lead placement.
2. Check/perform skin preparation.
3. Check/replace electrodes.
Check amplitude of ECG waveform:
1. Select ECG parameter label.
2. Select DISPLAY LEAD,
3. Scroll through all ECG leads and check for 0.5mV amplitude at normal (1X) size. (at least
0.5mV amplitude is required for QRS detection.) for borderline signals, validate on a graph.
4. If amplitudes are low, electrodes may need to be repositioned or replaced.
Problem :
False ventricular calls.
Solution :
Check ECG signal from Animal: (the chest lead may exhibit polarity changes which may
occasionally cause an inaccurate call.)
1. Check/adjust lead placement.
2. Check/perform skin preparation.
3. Check/replace electrodes. (if chest lead is a problem, move the chest lead to another chest
position or leg position.)
Problem :
Inaccurate pacemaker detection
Solution :
Use pacemaker processing:
1. Select ECG parameter label.
2. Display the lead of ECG with the greatest amplitude in the top waveform position.
3. Select ANALYSIS SETTINGS.
4. SELECT DETECT PACE.
Содержание BM3VET Touch
Страница 1: ...BM3VET Touch User s Manual Veterinary Patient Monitor Rev 1 1 20160401...
Страница 29: ...Rev 1 1 1 BASIC 29 BM3VET Touch User s Manual Features of Main Body FRONT BACK...
Страница 78: ...Rev 1 1 4 TREND 78 BM3VET Touch User s Manual 4 TREND 4 1 TREND GRAPHIC TREND TABLE TREND TREND WINDOW SETUP...
Страница 87: ...Rev 1 1 5 ECG 87 BM3VET Touch User s Manual IEC 3LEAD CABLE AHA 3LEAD CABLE IEC 5LEAD CABLE AHA 5LEAD CABLE...
Страница 88: ...Rev 1 1 5 ECG 88 BM3VET Touch User s Manual IEC 3LEAD AHA 3LEAD IEC 5LEAD AHA 5LEAD...
Страница 91: ...Rev 1 1 5 ECG 91 BM3VET Touch User s Manual 5 Position of 5 Lead RL V LL LA RA...
Страница 92: ...Rev 1 1 5 ECG 92 BM3VET Touch User s Manual Position of 3 Lead Wire Electrodes LL LA RA...