14. TROUBLE SHOOTING
182
BM3VET Touch User’s Manual
14.7 Power failure
The adapter connector Yes
is in bad condition
No
The output voltage of the Yes
adapter is lower than 18V
No
Repair the power B/D
“Admit” has been Yes
selected in the menu
No
The Digital B/D (dig) Yes
Battery has been lower than 3V
No
Repair the Dig B/D
Replace the adapter
Replace the adapter
Execute the “admit:” function
(No data will be stored during battery
discharge)
Replace the battery (3volt)