On
Point
™
Scope System
4.2 Sterilization of the OnPoint
™
CAUTION:
Do not sterilize the OnPoint 1.2mm Scope System or any
of its components. Refer to cleaning section for instructions.
WARNING:
RISK OF SHOCK. Do not autoclave. Do not soak or
immerse in disinfecting fluids.
Do not use or store disinfecting fluids above or on the OnPoint
Integrated Visualization System.
4.3 Lamp Life Monitoring and Replacement
Each time you turn on the OnPoint unit, the monitor will display the number
of hours of lamp life remaining:
Hours Remaining Display Format
Display Interval
51 or more
LAMP HOURS REMAINING #
5 seconds
26 - 50 LAMP HOURS REMAINING #
10 seconds
6 - 25 LAMP HOURS REMAINING #
20 seconds
1 - 5
REPLACE LAMP
30 seconds
0
REPLACE LAMP
120 seconds
1-877-304-3738
Field replacement of the lamp should never be attempted. For
service, contact your Biomet Microfixation representative.
31
4.4 Recommendations for the Disposal of the Device
For the disposal of the OnPoint unit, follow the guidelines in Directive
2002/96/EC Waste Electrical and Electronic Equipment (WEEE), or send the
equipment to Biomet Microfixation for proper disposal free of charge. For the
disposal of the single use endoscope, accessories, and procedure kit follow
the procedures for the disposal of bio-hazardous materials.
4.5 Fuse Replacement
1. Unplug the OnPoint power cord from the wall outlet to ensure that no
electrical power is present.
2. Use a standard screwdriver
to pry the fuse holder free
from the back of the unit
as shown.