C. The image is blurry.
1. If you are using a camera coupler, rotate the focus ring to bring
the image into focus.
2. The scope, coupler or camera optics may be dirty. Please refer
to the manuals that came with your camera and scope for proper
care and cleaning.
D. There is no light or light output is low.
1. If light output seems low, check the connection between the
light output on the camera and the scope for particles and foreign
objects.
2. If the output is still low or there is no light, your lamp may be bad.
Please contact your service representative.
See § 4.3 Lamp Replacement
3. If your lamp is good or is now replaced, and light output is now
still low, the light fibers in the camera cable may be damaged.
Please contact Biomet Microfixation for service.
Please copy your serial
number here for future reference: