5.0 Troubleshooting
Scenario
A: The unit will not operate. No fans blowing, with a blank screen.
1. Check all power connections to the OnPoint
™
unit. Check the
power cable connection to the wall outlet and the rear of the
OnPoint unit.
See § 3.2.2 Setup for Each Procedure
2. Ensure that the voltage-selector switch is in the appropriate
position.
See § 3.2.2 Setup for Each Procedure
3. Check fuses.
See § 4.5 Fuse Replacement
4. Check the building breaker controlling the outlet you are using.
You may need to consult the proper building administration to ensure
power to your outlet.
5. Should these measures fail, please contact Biomet Microfixation
for technical support and/or service.
B: The unit’s fans are blowing but the monitor is blank.
1. Ensure the camera cable connector is properly seated in the
receptacle on the front of the unit.
See § 3.2.2 Setup for Each Procedure
2. Check for light output. If there is none, replace the lamp.
See § 4.3 Lamp Replacement
3. Contact Biomet Microfixation for technical support and/or service.
Make a note of the camera you are using. If you have multiple
camera attachments, the support person may ask you to try another
camera. It may be necessary to send the camera you are using for
service along with the OnPoint unit.