25
Electromagnetic Compatibility
• The use of accessories and cables other
than those supplied may result in increased
emissions or decreased immunity
of the equipment or system
• This equipment should not be used adjacent
to or stacked upon other equipment
• Portable and mobile RF communications
equipment can adversely affect the operation
of Medical Electrical Equipment
• In the event this equipment interferes with
the operation of other equipment, or
experiences interference from other
equipment, to continue treatment it will be
necessary to move the Biomet
®
EBI Bone
Healing System away from the source
of the interference as indicated in table 4
The Biomet
®
EBI Bone Healing System must
emit electromagnetic energy in order to
perform its intended function. Nearby electronic
equipment may be affected.
The Biomet
®
EBI Bone Healing System is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Group 1
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Class B
Not Applicable
Not Applicable
Emissions Test
Compliance
Electromagnetic Environment - Guidance
The Biomet
®
EBI Bone Healing System is intended for use in the electromagnetic environment
specified below. The customer or the user of the Biomet
®
EBI Bone Healing System should assure
that it is used in such an environment.
Guidance and Manufactures Declaration – Electromagnetic Emissions
Table 1