Cleaning Instructions
Treatment Completion
Biomet
®
EBI Bone Healing System
The controller and AC Wall Adapter can be
cleaned by wiping with a damp cloth and mild
soap. Do not immerse in water or use solvents
or other cleaning agents. DO NOT MACHINE
WASH OR DRY
SFLX Treatment Coils
SFLX coils are not sterile. Clean with mild soap
and water by wiping with a damp cloth. Do not
immerse in any liquid.
*NOTE: Please make sure the controller is
disconnected from the AC Wall Outlet before
cleaning.
Biomet
®
EBI Bone Healing System
The maximum recommended therapeutic
treatment period is nine months (approximately
270 days).
When the Control Unit reaches system endpoint,
the signal is locked. “SYSTEM ENDPOINT CALL
BIOMET” will appear on the LCD display. The
display will then turn OFF. This message will
appear every time the Control Unit is turned ON.
At this point, the system should be disposed
of in an environmentally safe manner. (See
Disposal/Recycling on page 22).
*NOTE: If the prescribing physician feels that
treatment with the Biomet
®
Bone Healing
System is still necessary after the System
Endpoint has been reached, please contact
Biomet for instructions and procedure to
replace the expired controller.