28
Electromagnetic Compatibility
(Continued)
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people
Rated Maximum Output
Power of Transmitter
W
150 kHz to 80 MHz
80 MHz to 800 MHz
150 MHz to 2.5 GHz
d = 3.5 √ P
d = 3.5 √ P
.35
.35
d = 7 √ P
.01
1.1
1.1
.7
.1
3.5
3.5
2.21
1
11.06
11.06
7
10
35
35
70
22.13
100
Separation Distance (Meters) According to Frequency of Transmitter
The Biomet
®
EBI Bone Healing System is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the Biomet
®
EBI Bone
Healing System can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile communications equipment (transmitters) and the Biomet
®
EBI Bone Healing System as recommended below, according to the maximum power output of the
communications equipment.
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the Biomet
®
EBI Bone Healing System