To avoid the accidental reset of the daily time, the
RESET function has a three second delay. When
pressed for more than one second, the controller
will beep. To clear the time back to zero, continue
holding the button until a second beep is heard
(approximately three seconds). Release the
RESET/Backlight button. This will clear the time
back to 0:00.
Following completion of your daily treatment,
you should do the following:
1. Make sure the controller is off. If you
have completed 10 hours, the unit will
automatically shut off. With less than 10
hours of treatment, you will need to manually
turn the controller off.
2. Insert the AC Adapter charger into the
controller to recharge the battery for your
next treatment (see Step 1: BATTERY
CHARGING page 8).
The Biomet
®
EBI Bone Healing System may
be used at home or at work. Your patient’s
schedule and lifestyle will determine the best
time for using the system. Many people find it
convenient to treat while they are sleeping.
Troubleshooting System Messages
Allow up to one minute for the display message
to change, after taking corrective action.
• “RECHARGE BATTERY TREATMENT
HALTED” If this message appears, the
battery needs to be recharged. This message
will only appear during a treatment session.
In order to recharge the battery and
continue treatment, connect the AC Wall
Adapter and turn the controller on. Refer
to Step 4: (Page 10) “RECHARGING THE
BATTERY”
“Check Connectors See Manual”
This message, accompanied by audible beeps,
appears when the controller is not properly
connected to the SFLX Treatment Coil when
the controller is turned on. The controller will
turn itself off. Be sure to check all electrical
connections between the controller, link cable
and treatment coil. When the controller is
turned back on and the message continues,
check all cable connections again. If you need
assistance, you should call 1.973.299.9300
and ask to speak to a Patient Support
Representative. Outside the United States
contact you local EBI/Biomet Distributor.
”Check Coil See Manual”
This message, accompanied by three to
four audible beeps, appears when the SFLX
Treatment Coil is disconnected, damaged,
or inappropriately flexed. The message will
stay on the display until the controller turns
itself off. If you need assistance, you should
call 1.973.299.9300 and ask to speak to a
Patient Support Representative. Outside the
United States contact your local EBI/Biomet
Distributor.
RECHARGE BATTERY
TREATMENT HALTED
CHECK COIL
SEE MANUAL
CHECK CONNECTORS
SEE MANUAL