S12 Vet Monitor User’s Manual
51
D.2.1 General technical alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
XX
Communication
error
Med
Not clearable
XX measurement module failure or
communication failure.
Note: “XX” represents the module name, such as ECG, SpO2, IBP, TEMP, etc.
D.2.2 ECG Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
ECG Self-test
Error
Med
Not clearable
Board failure. Please contact the
manufacturer for repair.
ECG Leads Off
Med
Sound and light
can be cleared
All ECG leads fall off or ECG cables
are not connected. Please check the
connection of ECG electrode plates,
lead wires and cables.
ECG XX Off
Med
Sound and light
can be cleared
The electrode is not firmly connected
with the patient or falls off, causing the
corresponding ECG lead to fall off.
Please check the connection of ECG
electrode plates, lead wires and cables.
ECG YY
Polarized
Low
Sound and light
can be cleared
ECG electrode polarization or poor
contact. Please check the connection of
ECG electrodes plates.
ECG Learning
Prompt
/
Relearn is triggered manually or
automatically
ECG Cable
Incompatible
Med
Not clearable
Use non-factory cables. Please replace
the original cable.
ECG Cable Has
Expired
Med
Not clearable
ECG cable has expired. Please replace
the cable.