S12 Vet Monitor User’s Manual
12-11
procedure:
1.
Select the NIBP parameter area.
2.
Select
【
Setup
】
submenu
;
3.
Set
【
Assisted Venipuncture Pressure
】
.
4.
Select
【
Assisted Venipuncture
】
at the bottom of the menu.
5.
Puncture vein and draw blood sample.
6.
Select
【
NIBP Start/Stop
】
quick key or
【
Assisted Venipuncture
】
button to
manually deflate the cuff. When performing a venipuncture, observe the inflation pressure
and the remaining time of the venipuncture in the NIBP parameter area.
12.10.
NIBP Maintenance
12.10.1.
NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. The NIBP
leakage test should be performed once every two years or when you doubt the NIBP
measurements. The NIBP leakage test should be performed by qualified service personnel
only.
12.10.2.
NIBP Calibration
The NIBP accuracy test should be performed once every two years or when you
doubt the NIBP measurements. The NIBP accuracy test should be performed by qualified
service personnel only.
12.11.
NIBP Troubleshooting
For more information, see
D Alarm Information.