S12 Vet Monitor User’s Manual
16-3
16.3.1.3.
Install infusion tube set into DM module
Push the drip chamber into the slot of the drip chamber of the DM module, and clamp
the pipeline which is connected to the lower part of the drip chamber into the clamping
groove of DM module, as shown in the figure. The DM module is fixed to a suitable
position by means of supporting brackets or hanging ropes. Then, exhaust the pipeline to
ensure that the gas in the pipe set is exhausted and close the flow clamp tightly.
(1)
The slot of the drip chamber
(2)
Liquid stop clamp
(3)
Start/Stop DM button
(4)
Liquid stop clamp reset button
(5)
Liquid level indicator line of DM module
WARNING:
Ventilating operation can only be performed when the infusion is not performed
and the infusion tube is not connected to the patient.
16.3.1.4.
Configure related parameters
If the unit of mL/h needs to be adopted, can switch “Drops/min” to “mL/h” in the
menu of the monitor and set the conversion parameter between the number of drops and
mL.
CAUTION:
The drip rate corresponding to the 1 mL infusion volume must be entered
2
1
5
4
3