
S12 Vet Monitor User’s Manual
7-6
When the alarm volume is set to 0, the alarm tone will be turned off, and an
alarm audio off icon will appear on the screen.
When the alarm volume is set to 0, the setting of high level alarm volume is
invalid.
7.6.3.2.
Setting the minimum alarm volume
The minimum alarm volume determines the minimum alarm volume setting. The
steps are as follows:
1.
Select
【
Main Menu
】
quick key→from
【
System
】
column to select
【
Maintenance
】
→input maintenance password→Enter.
2.
Select
【
Alarm
】
submenu→
【
Sound
】
submenu.
3.
Select
【
Minimum Alarm Volume
】
.
NOTE
:
You can set the minimum alarm volume to 0 only when you are connected to
the CMS. If the monitor is not connected to the CMS, the minimum alarm
volume can only be set to 1.
When the CMS is connected, if the minimum alarm volume is set to 0, the
minimum alarm volume will be automatically changed to 2 when the CMS is
disconnected.
7.6.3.3.
Setting Alarm Tone Mode
The setting steps are as follows:
1.
Select
【
Main Menu
】
quick key→from
【
System
】
column to select
【
Maintenance
】
→input maintenance password→Enter.
2.
Select
【
Alarm
】
submenu→
【
Sound
】
submenu→
【
Alarm Sound
】
, you can
select
【
ISO
】
.
7.6.3.4.
Setting Alarm Tone Interval
You can set the alarm tone interval. The steps are as follows:
1.
Select
【
Main Menu
】
quick key→from
【
System
】
column to select