S12 Vet Monitor User’s Manual
22
positively bias Carbon Dioxide
values by up to an additional 3
mmHg at 38 mmHg.
Ethanol
0.1%
No Additional Effect
Isopropanol
0.1%
No Additional Effect
Acetone
0.1%
No Additional Effect
Methane
1%
No Additional Effect
Masimo ISA Capno & Masimo IRMA
Gas
Gas level
CO2
Agents
N2O
ISA CO2
ISA AX+
ISA OR+
N2O
4)
60 vol%
–
2)
–
1)
–
1)
–
1)
HAL
4)
4 vol%
–
1)
–
1)
–
1)
–
1)
ENF, ISO, SEV
4)
5 vol%
+8% of
reading
3)
–
1)
–
1)
–
1)
DES
4)
15 vol%
+12% of
reading
3)
–
1)
–
1)
–
1)
Xe (Xenon)
4)
80 vol%
−10% of reading
3)
–
1)
–
1)
He (Helium)
4)
50 vol%
−6% of reading
3)
–
1)
–
1)
Metered dose
inhaler
propellants
4)
Not for use with metered dose inhaler propellants
C2H5OH
(Ethanol)
4)
0.3 vol%
–
1)
–
1)
–
1)
–
1)
C3H7OH
(Isopropanol)
4)
0.5 vol%
–
1)
–
1)
–
1)
–
1)
CH3COCH3
(Acetone)
4)
1 vol%
–
1)
–
1)
–
1)
–
1)
CH4 (Methane)
4)
3 vol%
–
1)
–
1)
–
1)
–
1)
CO (Carbon
1 vol%
–
1)
–
1)
–
1)
–
1)