ONCORE PRO, Operating Manual, Rev. K
87
Section 8.
General Precautions
Make sure the slides are placed securely on the module –
the slide should be pushed in against the spring and the
spring should push the slide back under the slide clips.
Place the reagent rack firmly in its seated position
before starting a run on the instrument.
Remove caps from reagent vials
before starting a run on the instrument.
Keep the door closed during operation.
The robotic arm will move unexpectedly during operation - stay clear.
Do
not impair the movement of the robotic arm in any way.
Contact Customer Service prior to using reagents and solutions supplied by other vendors
on your Slide Stainer.
Some solvents, acids, and other solutions may cause damage to the internal components of the Slide Stainer and
affect your instrument's performance and warranty.
Wear disposable gloves and protective lab wear when handling reagents.
Reagents are harmful and irritating to
the eyes, respiratory system and skin. It may cause lung and stomach damage if ingested. SDS are available from
your local sales representative or online at biocare.net.
Hazardous reagent wastes must be disposed of according to local, state and federal regulations.
Wear
appropriate Personal Protective Equipment to prevent exposure.
Do not multitask or run CD/DVD players on the computer while the instrument is in operation.
Multitasking is
defined as running more than one software application at a time. This may lock up the instrument.
Disable Windows Automatic Update and other background programs
while the instrument is in operation.
Disable the Local/Wireless Internet Connection
. The system should not be connected to the Internet.
Do not modify Windows Power options or install a screen saver.
This may cause the computer to power off
unexpectedly while the instrument is in operation or may lock up the instrument.
Do not install third party software or hardware products.
Installing third party products may lock up the
instrument and may void the warranty.
Do not make any hardware or software changes
before consulting Technical Support.
Do not use a USB cable longer than 3 meters (10 feet).
Restart the computer if:
1)
The USB is disconnected/reconnected.
Always connect the USB to the designated USB port.
2)
The power supply to the instrument is powered on/off or disconnected.
3)
The computer is put into Sleep Mode or Hibernation Mode.
4)
The user encounters an unexpected error.
Do not attempt to service the Oncore Pro Slide Stainer
unless instructed to do so by Technical Support. Doing so
will void the warranty.
Do not relocate the Oncore Pro Slide Stainer System to another site within your facility
before contacting
Customer Service for vital information that may affect your warranty.