ONCORE PRO, Operating Manual, Rev. K
12
2.3.
Technical Specifications
Dimensions W x H x D
36.5in x 22in x 24in/
93cm x 56cm x 61cm
(with door closed)
36.5in x 36in x 24in/
93cm x 91cm x 61cm
(with door open)
Weight
125lb
(57kg)
Electrical requirements
120V 110/120V (±10%) 60Hz (±2Hz) 850 watts
220V 220/240V (±10%) 50Hz (±2Hz) 850 watts
Mains Connection: IEC320
Normal operating temperature
18°C-26°C
(64°F-79°F)
Slide capacity
1 - 36 glass slides
Reagent capacity
40 different reagents (15 mL/7mL reagent vial)
Heating capacity
Waste Separation
Room temperature to 103°C. Maximum 110°C.
Reagent dispense volumes
130uL standard test volume
(65 minimum-400 µl maximum)
1 hazardous waste, 1 non-hazardous waste
Reagent probe volume capacity
20 µL minimum-4500µL maximum
Computer controller requirements
BioCare reserves the right to change the computer controller specifications at any time.
Intel Core i5-2410M CPU 2.3GHz
4GB RAM
500GB Hard drive
4 USB 2.0 communication ports
Computer/Monitor
Laptop or Desktop Computer
Operating System
Windows 7, 10
Printer
Zebra TLP 3844-Z, Zebra GX430t or Zebra ZD620t
Surge Protector
120/240 VAC, 15 Amps, 50/60 Hz
LIS Connectivity
Compatible with SQL Database and HL7 messaging standards.
Operating Logic
Designed to calculate the most time-efficient sequence of steps to complete a programmed staining run.
Protocol Logic
Protocols are generated from a common protocol template optimized for each detection system.
End-users have the limited ability to customize key protocol steps.
Advanced users have full flexibility to add/remove and customize each step in the protocol template and to
generate special protocols.