2
Transportation, Storage and Handling
Storage:
environment is 10-40 °C (50-104 °F). Do not stack. Protect from rain. Duration not to exceed 1 year; may be extended with inspection
Transportation:
Shipping environment is 0-50 °C (32-122 °F). Do not stack. Protect from rain. In addition to the ASTM regulation
required markings, label shipping container with shipping and storage temperatures, a warning indicating do not stack and protect
from rain.
Handling:
The Decloaking Chamber may be safely handled by one person capable of lifting 20 pounds (18 kg). Lift by the handles
only. Do not move until completely cooled.
Action Upon Delivery
Unpacking should be performed or supervised by trained laboratory personnel or facility engineer who will record any issues with regard to
missing or damaged contents. In the event the instrument is damaged in transit, do not operate and contact Biocare Medical for instructions.
Operating Environment
Laboratory conditions: 15-30 °C; 15-80% relative humidity; maximum elevation 2000 M
Unpacking
Review and follow the instructions on the "Important Shipping Notice" insert.
Review the Contents of Your Package
Upon receipt of the Decloaking Chamber, inspect the packaging and become familiar with the parts. Save the packaging in case
returning the instrument becomes necessary.
Note
Be sure to remove any packing material from inside of the Decloaking Chamber before operating!
Failure to do so will result in FIRE.
Parts List
`
Lid with pressure gauge and hot sticker
`
Gasket (removable)
`
Weight (petcock)
`
Pan (removable)
`
Rack holder (removable) - only with Cat. No. DC2008US-RH and DC2008INTL-RH
`
Body with programmable digital timer
`
Electric power cord
`
Heat shield