1
Overview
The new and improved Decloaking Chamber Plus has been redesigned with a programmable walk-away capability, a cooling fan for
more rapid antigen retrieval and an optional stainless steel rack holder for plastic slide containers. Furthermore, a USB port enables the
laboratory to utilize a flash drive to track and save run data for quality control purposes. The pressure gauge has a new feature in which
the needle remains at the highest pressure reading during the run, validating that antigen retrieval has occurred properly. This feature
allows for a complete walk-away procedure. When the procedure is completely finished, an audible beep alerts the user.
The Decloaking Chamber is an excellent tool for heat-induced epitope-retrieval (HIER) methods. The proper use of heat and pressure
in conjunction with the appropriate buffer solutions is of the utmost importance for consistent immunohistochemistry (IHC) staining.
The Decloaking Chamber is designed to optimize and standardize antibody staining procedures and has been engineered to pass strict
laboratory safety requirements. Temperature, pressure and pH can be monitored and recorded with the Decloaking Chamber to produce
consistent staining, which will reduce repeat testing and false negatives.
Specifications
Decloaking Chamber - Intended Use
The Decloaking Chamber is a programmable bench top pressure cooker intended for laboratory use only. It is programmed to allow the
precise pressurized heating necessary for antigen retrieval and also has the capability to perform at a variety of temperatures ranging
from 37 °C to 125 °C + 5 °C.
Model: DC2008US
Model: DC2008US-RH
Voltage: 110-120 volts, 800 watts, 60 HZ
Model: DC2008INTL
Model: DC2008INTL-RH
Voltage: 220-240 volts, 900 watts, 50 HZ
Dimensions:
31cm (W) X 29cm (D) X 28cm (H)
Pan Capacity:
3.2 liters
Weight:
3.7 kg (8.18 lbs)
CE Listed
ETL Approved
START
STOP
DISPLAY
SET
Lid
Body
LCD
On/Off button
P1
SET HH:MM:SS
00:00:30
125.0
°