Continue to pump water through the system for at least 10 minutes.
You may wish to continue fraction collection or return the
SV3-2 valve to position 1 and then stop fraction collection.
Note
: In
manual
operation, starting or stopping fraction collection
does not change the SV3-2 diverter valve position. You must always
manually set the SV3-2 valve to the waste or collect position.
b. Using a Model 2128 Fraction Collector.
The Model 2128 Fraction Collector has two operating modes:
•
System: Controlled by the BioLogic HR System.
•
Local: Controlled from its own faceplate.
For now, insure that the System button is selected.
When in System mode, operation of the fraction collector is as follows.
•
Control of the Model 2128 is identical regardless of whether or not
the optional on-arm diverter valve is used. Without a valve, the
drop-head returns to the waste trough during divert operations.
•
The fraction collector box on the manual screen will show fields
for
Rack type
,
Start tube
,
End tube
,
Fraction size–ml
,
Tube
number
,
Volume left
, a toggle button for
Start
and
Stop
and a
button for
Advance
(see Figure 7).
Fig. 7. Manual Control Screen for the Automated BioLogic System.
BioLogicHR - Default - no method
File
View
Utilities
Options
Edit
Method
New
Method
New
Run
Browser
Manual
Setup Protocol
wash
load
1
2
Run
Notes
Advance
Gradient Pump
UV
0.00 ml/min
0 %B
438 psi
-0.00181 AU
Fraction Collector
START
Gradient pump
Inlet A:
1.00
100
0
Inlet B:
Flow rate:
ml/min
%
%
High
limit:
0
1000
Low
limit:
psi
AV7-3 at port 4
SV3-2 at port 1
SV5-4 at port 2
P
L
2
1
2
1
3
4
STOP
START
STOP
Set
psi
Tube number:
ml
Volume left:
Report
PostRun
Log
ON
Window
Help
Mode:
System
Local
Start Tube:
End Tube:
Fraction size:
1
1
128
1.00
Rack:
Rack 1 (128 tubes)
I
AV9-8 at port 5
3
7
8
2
6
5
4
I
Conductivity
47.5 mS/cm
0.089 Volt
7.00 pH
SIM2/pH
SIM1/SIG
Chart Recorder
Signal Import Module
SIM 1 and SIM 2 connected
UV Detector
Conductivity
range(mS/cm):
500
UV range (AU):
0.1
ON
OFF
Event mark
ON
OFF
Zero Baseline
Flowrate of the gradient pump
19