4. The Setup Editor is now complete and you are now ready to program
the separation steps. To see the protocol editor, click on the Protocol
Editor toolbar menu button.
Programming the Protocol Editor
1. From the Options pull-down menu, insure that Use Volume (ml) is
selected, so that the programming base is Volume.
Fig. 4. Protocol editor.
2. Program the separation method listed below and in Figure 4.
•
From the left side of the screen, press the
Fraction collection
button. In the pop-up window that appears, choose
Collect All
with a fraction size of 2.00 ml and a delay of 0. Make sure that you
choose the correct rack number for the rack you will be using
with the Model 2128.
•
Program the remaining steps using the add step icons from the
left-hand side of the screen.
Cut
Copy
Paste
Delete
Isocratic
Flow
Add Step
Linear
Gradient
Load/Inject
Sample
Change
Valve
Hold
Alarm
Zero
Baseline
UV Lamp
Chart
Recorder
Repeat
Steps
Fraction
Collection
Edit
View
Utilities
Options
Edit
Method
New
Method
New
Run
Manual
Setup Protocol
wash
load
1
2
Run
Notes
Report
PostRun
Log
Window
Help
1
2
3
6
5
0.0
0.0
0.0
1.5
Collect Fractions of size 2.00 ml during entire run
Chart Recorder
1.0
Turn ON
7
Isocratic Flow
4
1.0
Set Zero Baseline
Load/Inject Sample
UV Detector
2.3
Linear Gradient
8
9
15.3
18.1
Isocratic Flow
Isocratic Flow
Isocratic Flow
26.1
100%
100%
0%
0%
Load: Sample
Static Loop
25mM tris
25mM tris, 0.5M NaCl,
Manual Inject Valve
Flow: 4.00 ml/min
Volume: 2.8 ml
Flow: 4.00 ml/min
100%
0%
Volumn
Description
Parameters
100%
0%
Volume: 0.8 ml
Flow: 4.00 ml/min
Volume: 8.0 ml
Flow: 4.00 ml/min
25mM tris
25mM tris, 0.5M NaCl,
100% --> 50%
0% --> 50%
25mM tris
25mM tris, 0.5M NaCl,
25mM tris
25mM tris, 0.5M NaCl,
25mM tris
25mM tris, 0.5M NaCl,
Volume: 1.0 ml
Volume: 0.5 ml
Flow: 4.00 ml/min
Volume: 13.0 ml
Flow: 4.00 ml/min
End of Protocol
Gradient Pump: F-10
UV
0.00ml/min
0 %B
1 psi
0.00375 AU
Conductivity
3.59 mS/cm
SIM1/pH
SIM1/SIG
-0.003 Volt
File
Browser
Enter volume of sample to load
BioLogic Duo-Flow - Dr. J. Smith - Starter Kit - UNO Q1 -
11
4006147a.qxd 12/22/2000 1:29 PM Page 11