KBF-S / KBF-S-UL (E6) 11/2021
page 68/95
23. Maintenance and service, troubleshooting, repair, testing
23.1 General information, personnel qualification
•
Maintenance
See chap. 23.2
•
Simple troubleshooting
Chap. 23.4 describes troubleshooting by operating personnel. It does not require technical intervention
into the chamber, nor disassembly of chamber parts.
For personnel requirements please refer to chap. 1.1.
•
Detailed troubleshooting
If errors cannot be identified with simple troubleshooting, further troubleshooting must be performed by
BINDER Service or by BINDER qualified service partners or technicians, in accordance with the de-
scription in the Service Manual.
For personnel requirements please refer to the Service Manual.
•
Repair
Repair of the chamber can be performed by BINDER Service or by BINDER qualified service partners
or technicians, in accordance with the description in the Service Manual.
After maintenance, the chamber must be tested prior to resuming operation.
•
Electrical testing
To prevent the risk of electrical shock from the electrical equipment of the chamber, an annual repeat
inspection as well as a test prior to initial startup and prior to resuming operation after maintenance or
repair, are required. This test must meet the requirements of the competent public authorities. We rec-
ommend testing under DIN VDE 0701-0702:2008 in accordance with the details in the Service Manual.
For personnel requirements please refer to the Service Manual.
23.2 Maintenance intervals, service
DANGER
Electrical hazard during live maintenance work.
Deadly electric shock.
∅
The chamber must NOT become wet during operation or maintenance works.
∅
Do NOT remove the rear panel of the chamber.
Disconnect the chamber before conducting maintenance work. Turn off the main
power switch and pull the power plug.
Make sure that general maintenance work will be conducted by licensed electricians
or experts authorized by BINDER.
Make sure that maintenance work at the refrigeration system will only be conducted by
qualified personnel who underwent training in accordance with EN 13313:2010 (e.g. a
refrigeration technician with certified expert knowledge acc. to Regulation (EC) nº
303/2008). Follow the national statutory regulations.
Ensure regular maintenance work is performed at least once a year and that the legal requirements are
met regarding the qualifications of service personnel, scope of testing and documentation. All work on the
refrigeration system (repairs, inspections) must be documented.
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...