KBF-S / KBF-S-UL (E6) 11/2021
page 16/95
1.11 Operating instructions
Depending on the application and location of the chamber, the operator of the chamber must provide the
relevant information for safe operation of the chamber in a set of operating instructions.
Keep these operating instructions with the chamber at all times in a place where they are
clearly visible. They must be comprehensible and written in the language of the employees.
1.12 Measures to prevent accidents
The operator of the chamber must observe the local and national regulations (for Germany: the rule “Op-
eration of work equipment. Operation of refrigeration systems, heat pumps and refrigeration equipment”,
GUV-R 500 chap. 2.35) and take precautions to prevent accidents.
The manufacturer took the following measures to prevent ignition and explosions:
•
Indications on the type plate
See operating manual chap. 1.6.
•
Operating manual
An operating manual is available for each chamber.
•
Overtemperature monitoring
The chamber is equipped with a temperature display, which can be read from outside.
The chamber is equipped with an additional safety controller (temperature safety device class 3.1 acc.
to DIN 12880:2007). Visual and audible (buzzer) signals indicate temperature exceeding.
•
Safety, measurement, and control equipment
The safety, measuring, and control equipment is easily accessible.
•
Electrostatic charge
The interior parts are grounded.
•
Non-ionizing radiation
Non-ionizing radiation is not intentionally produced, but released only for technical reasons by electrical
equipment (e.g. electric motors, power cables, solenoids). The machine has no permanent magnets. If
persons with active implants (e.g. pacemakers, defibrillators) keep a safe distance (distance of field
source to implant) of 30 cm, an influence of these implants can be excluded with high probability.
•
Protection against touchable surfaces
Tested according to EN ISO 13732-1:2008.
•
Floors
See operating manual chap. 3.4 for correct installation
•
Cleaning
See operating manual chap. 23.4.
•
Examinations
The chamber has been inspected by the “Deutsche Gesetzliche Unfallversicherung e.V. (DGUV) (Ger-
man Social Accident Insurance (DGUV)” (German Social Accident Insurance (DGUV), Testing and Cer-
tification Body for Foodstuffs and Packaging Industry in DGUV Test) and bears the GS mark. Not valid
for UL chambers.
UL chambers: The chamber is certified by Underwriters Laboratories Inc.
®
according to the following
standards: UL 61010-1, 3
rd
Edition, 2012-05, Rev. 2015-07, CAN/CSA-C22.2 No. 61010-1, 3
rd
Edition,
2012-05, Rev. 2015-07.
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...