KBF-S / KBF-S-UL (E6) 11/2021
page 55/95
Principle of entry:
•
Use the
OK button
to select the desired section of the DNS server address 1/4, 2/4, 3/4, 4/4 in the
upper display line
•
Use the
Arrow buttons
to enter the value for the selected section of the DNS server address
•
For details please refer to the description of the similar procedure in chap. 15.2.3 “Assigning the IP
address”.
With the
Back button
you can go back to the “
Ethernet
” submenu and, repeatedly pressing it, to
Normal
display
.
16. Data recorder
An internal data recorder saves chamber data and events in three data sets.
With the export function “Export recorder data” (chap. 17.3) you can save the three data sets via the USB
interface to USB stick. They are issued in the selected language as a spreadsheet with the file extension
“.csv” and can be further processed in the desired program. The data is unencrypted. Always the entire
data memory is read out.
The USB stick must be formatted with FAT32 and have at least 8GB of memory.
16.1 Recorded data
All data is given out in tabular form. The headlines of the values “number”, “date”, and “time” are given out
in the selected language, all other information in English.
•
Chamber data for the user “DL1”
Tabular representation of the actual temperature and humidity values together with the date and time,
according to the set storage rate (chap. 16.3). Temperature values are always given out in °C.
•
Chamber data for BINDER Service “DL2”
This data is intended for use by BINDER Service. It also contains information from the self-test function.
The storage rate is fix (1 minute). Temperature values are always given out in °C.
•
Event list
Messages regarding the controller and data memory as well as the alarm messages together with the
date and time:
•
Firmware update done
•
“New config (USB)”: New configuration uploaded via USB
•
“Data recorder cleared”: Data recorder and event list deleted via setup program
•
Other event messages according to existing alarms
•
The moment of switching the alarm state on and off is indicated under “On/Off”.
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...