KBF-S / KBF-S-UL (E6) 11/2021
page 24/95
DANGER
Electrical hazard by contact of electrical wiring or components with water.
Deadly electric shock.
∅
Do NOT conduct any electrical wiring below the chamber.
∅
Do NOT install any electrical components under the chamber.
Ambient conditions
•
Permissible ambient temperature range during operation: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C /
71.6 °F +/- 5.4 °F
to which the specified technical data relate.
Deviations from the indicated data are possible for other ambient conditions.
With each degree of ambient temperature >25 °C /
77 °F
, the refrigeration power decreases
by 1.5 K.
•
Permissible ambient humidity: 70 % r.h. max., non-condensing
When operating the chamber at temperature set-points below ambient temperature, high ambient humidity
may lead to condensation on the chamber.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
Minimum distances
•
When placing several chambers of the same size side by side, maintain a minimum distance of 250 mm
/
9.84 in
between each chamber.
•
Wall distances: rear 100 mm /
3.9 in
, sides 160 mm /
6.29 in
.
•
Spacing above the chamber of at least 100 mm /
3.9 in
must also be accounted for.
Maximum load of the top of the housing: 15 kg /
33 lb
. The chambers are NOT intended for stacking.
NOTICE
Danger by stacking.
Damage to the chambers.
∅
Do NOT place the chambers on top of each other.
Other requirements
Water supply is provided by manually filling the freshwater can (chap. 4.1).
To avoid any possible water damage, provide a floor drain at the location of the device. Se-
lect a suitable installation site to avoid any consequential damage by splashing water.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing) several
times a year.
Avoid any conductive dust in the ambiance according to the chamber layout complying with pollution degree
2 (IEC 61010-1).
Install the chamber in a way that freshwater can is easily accessible for filling
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...