KBF-S / KBF-S-UL (E6) 11/2021
page 33/95
Functional controller keys
Icon
Signification
Function
Arrow-up button
•
Navigate between menus, submenus, other functions
•
In the setting menu: change setting, decrease value
Arrow-down button
•
Navigate between menus, submenus, other functions
•
In the setting menu: change setting, increase value
OK button
•
Select menu, submenu, function
•
In the setting menu: Confirm entry
Back button
Back to previous menu level
Standby button
no function
5.1 Menu structure of the controller and access levels
Starting from
Normal display
, navigate between the menus with the
arrow buttons
.
With the
OK button
you enter the setting of further subordinate menu functions.
With the
Back button
you go back to the previous function and finally back to Normal display.
The available functions depend on the current dependent on the current
authorization
“User”, “Admin” or
“Service”, for which the entry of a password may be required, depending on the setting.
You can set passwords for different access levels:
•
User:
The password enables access to the standard operating functions. Factory setting is 00 00 (no
password assigned).
•
Admin:
The password enables access to advanced controller functions and settings. Factory setting is
00 01.
•
Service:
The password enables access to all controller functions (for BINDER Service only).
As soon as a password has been assigned, access to the respective functions is blocked and only available
after entering the correct password.
Menu
Required access level
Functions
Setpoints
“User”
•
Temperature and humidity set-point setting
•
Setting the safety controller
•
Turning on / off humidity control
•
Activating/deactivating operating mode “Idle mode”
Chamber info
Any user
•
Configuration display (setup information, controller hard-
ware and software, analog inputs)
•
Display of interface configuration (e.g. MAC address, IP
address)
Settings
“Admin”
•
General controller settings (date, time, menu language,
temperature unit, display brightness…)
•
Network settings
•
Setting the data logger storage interval
•
Setting the tolerance range limits and delay time for toler-
ance range alarm
•
Self-test function
•
Password changing for User and Admin
Содержание 9020-0366
Страница 81: ...KBF S KBF S UL E6 11 2021 page 81 95 25 7 Dimensions Dimensions size 115...
Страница 82: ...KBF S KBF S UL E6 11 2021 page 82 95 Dimensions size 240...
Страница 83: ...KBF S KBF S UL E6 11 2021 page 83 95 Dimensions size 720...
Страница 84: ...KBF S KBF S UL E6 11 2021 page 84 95 Dimensions size 1020...
Страница 86: ...KBF S KBF S UL E6 11 2021 page 86 95...
Страница 87: ...KBF S KBF S UL E6 11 2021 page 87 95...
Страница 89: ...KBF S KBF S UL E6 11 2021 page 89 95...