Calibrating and adjusting the temperature and humidity control
Use the buttons
or
to select a menu item and confirm
with <ENTER>. For each menu item a submenu will be
opened.
Use the buttons
,
,
and
to key in the values de-
scribed below and confirm with <ENTER>.
State: On
Start value X: 0.0
Start value W: 0.0
End value X: Value of “Safety controller“/“1.” (e.g.
+22.7) (step 0, chap. 11.3.2)
End value W: Value of reference measuring in-
strument (e.g. +21.6) (step 0, chap. 11.3.2).
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue input 1
Analogue input 2
Analogue input 3
Analogue input 4
Analogue input 5
Safety controller
Safety controller
State
On
Start Value X
+
0.0
Start Value W
+
0.0
End Value X
+
22.7
End Value W
+
21.6
↓
Push <EXIT> four times to return to the basic display.
The corrected value appears on the controller display
after 5 seconds.
Compare the displayed temperature values of the
controller display and the reference measuring
instrument. The deviation must be less than the
confidence criteria.
W
X
128
Service Manual KBF (E5.2) 04-2015