Calibrating and adjusting the temperature and humidity control
Use the buttons
or
to select a menu item and confirm
with <ENTER>.
For each menu item a submenu will be opened.
Use the buttons
,
,
and
to key in the values given
below and confirm with <ENTER>.
State: Off
Start value X: 0.0
Start value W: 0.0
End value X: 1000.0
End value W: 1000.0
Configuration 2
Analogue inputs
Analogue outputs
Limit comperators
Donor-functions
Controller
Service Level
Service Level
Inp.Val. correct.
Analogue input 1
Analogue input 2
Analogue input 3
Analogue input 4
Analogue input 5
Safety controller
Safety controller
State
Off
Start Value X
+
0.0
Start Value W
+
0.0
End Value X
+1000.0
End Value W
+1000.0
↓
Push <EXIT> four times to return to the controller’s
normal display.
W
X
124
Service Manual KBF (E5.2) 04-2015