Microtitre Plate 
Shaker Incubator

SI505

Instructions for use

Introduction

Thank you for purchasing this piece of Stuart equipment. 
To get the best performance from the equipment and for 
your own safety, please read these instructions carefully 
before use. Before discarding the packaging check that 
all parts are present and correct.

This equipment is designed to operate under the 
following conditions:

• 

For indoor use only

• 

Use in a well ventilated area

• 

Ambient temperature range +10ºC to +40ºC

• 

Altitude to 2000m

• 

Relative humidity not exceeding 80%

• 

Mains supply fluctuation not exceeding 10%

• 

Over-voltage category II IEC60364-4-443

• 

Pollution degree 2

If the equipment is not used in the manner described 
in this manual and with accessories other than those 
recommended by Stuart, the protection provided may 
be impaired.

Electrical Installation

      THIS EQUIPMENT MUST BE EARTHED

Before connection please ensure that the line 
supply corresponds to that shown on the rating 
plate. 

Model 

Supply requirements 

Power

SI505 

230V, 50 Hz, 

250W

The model is supplied with two mains leads fitted with 

IEC plugs for connection to the instrument. One has 
a U.K. 3 pin plug and the other has a 2 pin “Shuko” 
plug for connection to the mains. Choose the lead 

appropriate for your electrical installation and discard 
the other. Should neither lead be suitable, take the lead 
with the U.K. plug and replace the plug with a suitable 
alternative. This involves cutting off the moulded plug, 

preparing the cable and connecting to the re-wireable 

plug in accordance with its instructions.

THIS OPERATION SHOULD ONLY BE UNDERTAKEN 
BY A QUALIFIED ELECTRICIAN

NOTE

: Refer to the equipment’s rating plate to ensure 

that the plug and fusing are suitable for the voltage and 
wattage stated 

The wires in the mains cable are coloured as follows:

BROWN 

 

-  

LIVE

BLUE  

 

-  

NEUTRAL

GREEN/YELLOW  

-  

EARTH

The instrument is fitted with an IEC socket at the rear 

of the instrument for connection of the mains lead. The 
appropriate mains lead should be connected 

BEFORE

 

connection to the mains supply. 

Should the mains lead need replacement, a cable of 

1mm2 of harmonised code H05W-F connected to an 
IEC320 plug should be used. 

IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN

Safety Advice Before Use

• 

Do not attempt to lift or carry the unit alone.

• 

Never move or carry the unit while it is in use or  

 

connected to the mains supply.

• 

Do not use in a hazardous atmosphere or with  

 

hazardous materials.

• 

Ensure that the load is balanced.

• 

Do not attempt to stop movement by hand.

• 

In the case of mains interruption, the unit will  

 

not restart on restoration of the electricity supply.

• 

In the case of mechanical interruption (e.g.  

 

motor stall), the unit will continue operating on  

 

removal of interruption.

Some parts of the incubator may become hot 
during use. These are clearly marked with 
warning labels and include the heated base and 

steel parts of the water reservoirs. Avoid touching 

these parts when the incubator is running.

General Description

The Stuart SI505 microtitre plate shaker incubator is 

ideal for mixing and aerating samples in microtitre 

plates or micro centrifuge tubes under accurate 

temperature controlled conditions. High speed 

combined with a tiny orbit creates the vibrational 
shaking action required for very small sample volumes. 

The retractable platform allows for easy access when 

loading or unloading samples. This has a non-slip mat 

to hold microtitre plates in place and micro centrifuge 
tubes of various sizes can be held in purpose built 
accessory racks.

The incubator chamber is fitted with two water 
reservoirs which allow the humidity of the chamber to 
be increased, thus reducing the possibility 

of sample 

loss due to evaporation over longer incubation 

periods.

Содержание Stuart SI505

Страница 1: ...505 Instructions for use Manuel d utilisation Istruzioni per l uso Instrucciones de funcionamiento Bedienungsanweisung Version 1 0 press to set on off temperature C on off time rpm start stop mode mic...

Страница 2: ...otitre plate skaker incubator S1505 microtitre plate skaker incubator S1505 press to set on off temperature C on off time rpm start stop mode Temperature control panel USB port Speed time control pane...

Страница 3: ...s in the mains cable are coloured as follows BROWN LIVE BLUE NEUTRAL GREEN YELLOW EARTH The instrument is fitted with an IEC socket at the rear of the instrument for connection of the mains lead The a...

Страница 4: ...of water may be found on the bottom of the chamber This should not affect the safety or performance of the product but regular drying with a cloth is recommended Loading the samples 9 Lift the release...

Страница 5: ...n 29 Click Run Software to complete the program start up Maintenance Servicing WARNING Ensure the unit is disconnected from the mains electricity supply before attempting maintenance or servicing Peri...

Страница 6: ...0 1 C Temperature fluctuation 0 5 C at 37 C Speed range 250 to 1250rpm Orbit diameter 1 5mm Maximum load 1kg Speed control Digital set in 10rpm increments Timer min s h min day h Electrical supply 23...

Страница 7: ...incubator S1505 microtitre plate skaker incubator S1505 press to set on off temperature C on off time rpm start stop mode Panneau de commande de la temp rature Port USB Panneau de commande de la vite...

Страница 8: ...ance mentionn es Les couleurs des fils du cordon d alimentation sont les suivantes MARRON PHASE BLEU NEUTRE VERT JAUNE TERRE L appareil est quip d une prise IEC sur sa face arri re pour la connexion d...

Страница 9: ...e fa on les extraire totalement ou partiellement de l appareil 7 Remplissez chaque r servoir avec de l eau distill e ou d ionis e jusqu au rep re de remplissage maximum Ce volume est normalement suffi...

Страница 10: ...p Appuyez de nouveau sur le bouton start stop pour continuer 23 Il est possible de r gler le minuteur tout moment sans arr ter l appareil en tournant le bouton de commande 24 Lorsque le compte rebours...

Страница 11: ...t des dispositifs de s curit int gr s l appareil En cas de doute contactez le service d assistance de Bibby Scientific Ltd ou votre revendeur Pour toute autre question technique contactez le service d...

Страница 12: ...ncubator S1505 microtitre plate skaker incubator S1505 press to set on off temperature C on off time rpm start stop mode Pannello di controllo della temperatura Porta USB Pannello di controllo della v...

Страница 13: ...pina e i fusibili siano adatti alla tensione e alla potenza indicate consultando la targhetta dei dati nominali dell apparecchio Colori dei conduttori nel cavo di alimentazione MARRONE FASE BLU NEUTRO...

Страница 14: ...fino al segno max fill livello massimo di riempimento con acqua distillata o deionizzata In tal modo l umidit della camera dovrebbe mantenersi intorno all 80 90 per 72 ore a 37 C 8 Ricollocare accurat...

Страница 15: ...le acustico Nota se lo sportello viene aperto mentre la piattaforma in movimento l apparecchio emette un segnale acustico e il conto alla rovescia si arresta Inoltre sul display viene visualizzata tem...

Страница 16: ...menti modifiche montaggio di componenti non approvati o riparazioni condotte da personale non autorizzato Bibby Scientific Ltd Beacon Road Stone Staffordshire ST15 0SA Regno Unito Tel 44 0 1785 812121...

Страница 17: ...e skaker incubator S1505 microtitre plate skaker incubator S1505 press to set on off temperature C on off time rpm start stop mode Panneau de commande de la temp rature Puerto USB Panel de control de...

Страница 18: ...ICISTA CUALIFICADO NOTA Consulte la placa de caracter sticas del equipo para asegurarse de que utiliza el enchufe y el fusible adecuados para la tensi n y potencia indicadas Los hilos del cable de ali...

Страница 19: ...ra 6 Abra la puerta y saque completa o parcialmente los tanques de agua situados en la parte superior de la c mara 7 Llene cada tanque hasta la marca de llenado m ximo con agua destilada o desionizada...

Страница 20: ...brevemente en la pantalla El calentador permanecer encendido ON Cierre la puerta para continuar y se reanudar la cuenta atr s en el punto en el que se detuvo Conectividad USB La unidad SI505 tiene la...

Страница 21: ...da o causado por una instalaci n adaptaci n modificaci n instalaci n de piezas no aprobadas o reparaci n llevadas a cabo por personal no autorizado Bibby Scientific Ltd Beacon Road Stone Staffordshire...

Страница 22: ...aker incubator S1505 microtitre plate skaker incubator S1505 press to set on off temperature C on off time rpm start stop mode Bedienfeld zur Regelung der Temperatur USB Schnittstelle Bedienfeld zur R...

Страница 23: ...eiden das Kabel vorbereiten und den verdrahtbaren Stecker gem den Anweisungen anschlie en DIESER VORGANG DARF NUR VON EINEM QUALIFIZIERTEN ELEKTRIKER DURCHGEF HRT WERDEN HINWEIS Pr fen Sie anhand des...

Страница 24: ...Ger t l sst sich jedoch nur auf mindestens 7 C oberhalb der Umgebungstemperatur einregeln d h bei einer Umgebungstemperatur von beispielsweise 10 C betr gt die Mindestsolltemperatur 17 C Erh hung der...

Страница 25: ...it w hrend des Betriebs anhalten indem Sie die Taste start stop dr cken Um den Betrieb fortzusetzen dr cken Sie erneut start stop 23 Der Timer l sst sich jederzeit ber den Reglerknopf einstellen ohne...

Страница 26: ...Vertretern geliefert oder angegeben werden Die Verwendung nicht genehmigter Teile kann die Leistung der in das Ger t integrierten Sicherheitseinrichtungen beeintr chtigen Bei Fragen oder Zweifeln wend...

Страница 27: ...Notes...

Страница 28: ...ications We cannot be sure that other equipment used in its vicinity will meet these standards and so we cannot guarantee that interference will not occur in practice Where there is a possibility that...

Страница 29: ...25 9243 Fax 609 589 2571 www bibby scientific com Bibby Scientific Italia Srl Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel 39 0 2 98230679 Fax 39 0 2 98230211 e mail mark...

Отзывы: