9.  IMPORTANT INFORMATION REGARDING ELECTROMAGNETIC COMPATIBILITY (EMC)

-  The use of accessories, transducers and cables other than those mentioned or provided by 

With the increased number of electronic devices such as computers and mobile (cellular) 
telephones, medical devices in use may be susceptible to electromagnetic interference from 
other devices. Electromagnetic interference may result in incorrect operation of the medical 
device and create a potentially unsafe situation. Medical devices shouldalso not interfere with 
other devices.

In order to regulate the requirements for EMC (Electromagnetic Compatibility) with the aim to 
prevent unsafe product situations, the IEC60601-1-2 standard has been implemented. This 

levels of electromagnetic emissions for medical devices.

The stimulator conform to this IEC60601-1-2:2007 standard for both immunity and emissions.

Nevertheless, special precautions need to be observed:
- The use of accessories other than those provided by manufacturer , may result in increased 
   emission or decreased immunity of the device.
- Refer to EMC table guidance regarding the EMC environment in which the device 
  should be used.

the manufacturer of this equipment could result in increased electromagnetic emissions or 
decreased electromagnetic immunity of this equipment and increase the risk of malfunction

8. TECHNICAL CHARACTERISTICS

Units

2 Units (49.9mmx55.8mm)

Programmes

19 programmes (11 TENS + 8 EMS programmes)

Battery

3.7V/500mAh  Li-lon

Output voltage

max 60mA  ±10% (1000Ω load)

Conditions of use

5°C to 40°C with a relative humidity of 15% to 93%, 
atmospheric pressure of 700 to 1060 hPa

Storage and Transportation Conditions  

-10°C to 55°C with a relative humidity of 10% to 
90%, atmospheric pressure of 700 to 1060 hPa

Dimensions

440x54x13mm cable included

Weight 

55g (±20%) by unit

IP22

Wave form

Biphasic square wave (net current 0dc) 

Impulse width

150us~370us (±10%)

Heart rhythm

2Hz~100Hz  (±10%)

Treatment time

22min/30min given by the application

Adapter

1pc (Input : 100-240 V AC 50/60Hz Output : 5 V DC 

Service life of the device

2 years 

Service life of electrode pads 

Applied part

 Electrode size

Eelctrode impadence 

10 times

Eelctrode

28*25mm(ellipse)

≤500Ω

300mA)

 

 

Guidance and manufacturer’s declaration – electromagnetic emissions

The customer or the user of these electrical stimulators should assure that it is used 
in such environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions  
CISPR 11

Group 1

The device uses RF energy only for its internal 
function. Therefore, its RF  emissions are very low 
and are not likely to cause any interference in 
nearby electronic equipment.  

RF emissions  

CISPR 11

Class B

The device is suitable for use in all 
establishments including domestic and those 
directly connected to the public low-voltage 
power supply network that supplies 
buildings used for domestic purposes. 

Harmonic emissions

 

lEC 61000-3-2

Class A

Voltage fluctuations/

 

flicker emissions 

 

lEC 61000-3-3

Complies

TABLE 1: 

Содержание Mytens BW-TSX

Страница 1: ...ser Manual 072015 1 1 BW TSX 0197 Manufacture Nexmed Technology Co LTD Address 2nd Floor of No 1 building Jia An Technological Industrial Park 67 District Bao An 518101 Shenzhen China BW TSX Web site...

Страница 2: ...e in case of anomalies or malfunction Electronic medical equipment requires special precautions regarding electromechanical com patibility It must be installed and used in accord with EMC ElectroMagne...

Страница 3: ...s Do not place the electrodes near cancerous lesions because this may have a negative impact on these injuries Do not place the electrodes on skin areas whose sensations are not normal You may burn yo...

Страница 4: ...ions Always consult your doctor or midwife if you are pregnant and intend to use the device cal muscle stimulation EMS device could in this case cause a blood clot In case of cognitive impairment If y...

Страница 5: ...Strengthening of the muscular mass EMS 8 standard programmes Stimulation of the venous return TENS Easy to use Data transfer onto a phone or a Bluetooth 4 0 tablet 4 USE 4 1 DE See schema in page 8 SC...

Страница 6: ...mage do not use the device or one of the electrodes Usable indoors and outdoors the device is not designed to withstand all weather conditions The device is not waterproof Do not wet it and never use...

Страница 7: ...angerous However if you experience discomfort it is advisable to decrease the intensity of the programme change or move the electrodes The burst transcuta neous electrical neurostimulation Burst TENS...

Страница 8: ...tion must remain comfortable and muscle contractions must not are advised to choose a low intensity to limit muscular fatigue SummaryTable of the EMS Programmes EMS Impulses Treatment site Name Treatm...

Страница 9: ...harge the device fully before beginning a session If the charge is not complete at the beginning of a programme the battery may become deplete before the end of the session You cannot use the device w...

Страница 10: ...g or biting sensation Note Always connect the device before you place the electrodes on the skin Note Do not place the electrodes on top of each other or so close to each other that they touch each ot...

Страница 11: ...ensity lever you want Starting and stopping a treatment Tap the play button on the screen The treatment screen opens and the time starts to count down The status indicator on the device starts to flas...

Страница 12: ...en stimulation is uncomfortable i e when you experience an unpleasant stinging or biting sensation Note Always replace the electrodes with electrodes recommended for this device by the manufacturer 5...

Страница 13: ...9mmx55 8mm Programmes 19 programmes 11 TENS 8 EMS programmes Battery 3 7V 500mAh Li lon Output voltage max 60mA 10 1000 load Conditions of use 5 C to 40 C with a relative humidity of 15 to 93 atmosph...

Страница 14: ...on electromagnetic IMMUNITY The customer or the user of these electrical stimulators should assure that it is used in such environment Immunity test IEC 60601 test level Compliance level Electromagnet...

Страница 15: ...ith the legal obligation For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequ...

Страница 16: ...Pain in the lower back Elbow Pain in the elbow Fore arm Pain in the forearm Wrist Pain in the wrist and carpal tunnel syndrome Hip Pain in the hip 12 Thigh Sciatic pain Knee Knee pain Lower leg Pain...

Страница 17: ...ard will not be extended to another per including the user This guarantee covers only the defects of the material or parts occurring during normal use of the pro duct It does not cover the damage caus...

Отзывы: