19
Problem/
Question
Possible Cause/Remedy
Which medi-
cines are suit-
able for inhala-
tion?
Only the doctor can advise you
which medicine to use to treat your
condition.
Consult your doctor.
With the IH 50, you can atomise
medicines with a viscosity lower
than 3.
Some inhala-
tion solution
remains in the
nebuliser.
This is normal and occurs for techni-
cal reasons. Stop inhaling as soon
as you can hear a marked differ-
ence in the sound made by the at-
omiser, or when the device switches
off automatically due to insufficient
inhalant.
What should be
taken into ac-
count when us-
ing the device
with infants and
children?
1. In babies, the mask should cover
the mouth and nose to guarantee
effective inhalation.
2. In children, the mask should also
cover the mouth and nose. Nebu-
lisation next to a sleeping person
is not suitable because insufficient
medicine reaches the lungs.
Note:
Children should only use the
device with help and under supervi-
sion of an adult. Never leave a child
alone with the nebuliser.
Inhalation with
the mask takes
longer.
This is for technical reasons. Less
medicine is inhaled per breath
through the mask holes than us-
ing the mouthpiece. The aerosol is
mixed with ambient air through the
holes in the mask.
Does each user
need their own
accessories?
This is absolutely necessary for hy-
giene reasons.
9. Technical specifications
Dimensions
(LxWxH)
72 x 59 x 143 mm
Weight
238 g incl. batteries
Battery operation
4 x 1.5 V type AA, Mignon (LR6)
Volume
max. 8 ml
Medicine flow
approx. 0.25 ml/min.
Oscillation
frequency
100 kHz
Housing material ABS / TPE
Mains
connection
100 – 240 V~; 50 – 60 Hz; 0,15 A
Operating
conditions
Temperature: +5 °C – +40 °C
Relative humidity: <80 % non-con-
densing
Storage and
transport
conditions
Temperature: -10 °C – +45 °C
Relative humidity: <80 % non-con-
densing
Subject to technical modifications.
Wearing parts are not covered by the guarantee.
The serial number is located on the device or in the
battery compartment.
All measurements were obtained with a sodium chlo-
ride solution using a laser diffraction method.
This diagram may not be applicable for suspensions or
highly viscous medicines. More information can be ob-
tained from the relevant medicine manufacturer.
10. Replacement parts and wear-
ing parts
Replacement parts and wearing parts are available
from the corresponding listed service address under
the stated material number.
Designation
Material
REF
Mesh atomiser with
medicine container
PC
162.711
Yearpack
(Mouthpiece, Adult mask,
Child mask, Mesh atomiser,
Connecting piece)
PP,
PVC, EVA,
PET
603.05
Note
If the unit is used outside of the specifications, proper
function is no longer guaranteed! We reserve the right
to make technical changes to improve and further de-
velop the product.
This device and its accessories comply with European
standards EN60601-1 and EN60601-1-2, as well as
EN13544-1, and is subject to special safety measures
in terms of electromagnetic tolerance. Note that port-
able and mobile RF communication equipment can af-
fect this unit. More details can be requested from the
stated Customer Service address or found at the end
of the instructions for use. The unit conforms to the re-
quirements of the European Directive for Medical Prod-
ucts 93/42/EC, the MPG (German Medical units Act).