Appendix V
I
Instructions for use
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.
Warning
:
Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging,
where the intensity of EM disturbances is high.
Warning
:
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper
operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating
normally.
Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper
operation.”
Warning
:
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be
used no closer than 30 cm (12 inches) to any part of the Digital Automatic Blood Pressure Monitor (model: MD6000), including
cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
If any
:
a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are
replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM
with the requirements of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either
generically (e.g. shielded cable, load impedance) or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE REFERENCE).
If any
:
the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a
description of what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM
DISTURBANCES (the defined term “ESSENTIAL PERFORMANCE” need not be used).