background image

Measur

ement

• If the 

Bluetooth

®

 function has been activated, the data transfer to the “beurer HealthManager” app 

starts automatically after the measurement.

• The 

Bluetooth

®

 symbol flashes on the display. The device now attempts to connect to the app for 

approx. 30 seconds.

• The 

Bluetooth

®

 symbol stops flashing as soon as the connection is established. 

All measurement data 

is transferred to the apps.  Once the data transfer is successfully complete, the device switches off 

automatically.

• If a connection to the app cannot be established after 30 seconds, the 

Bluetooth

®

 symbol goes out 

and the device switches off automatically after 1 minute.

• If you forget to turn off the device, it switches off automatically after approximately 1 minute.

 In this 

case too, the value is stored in the selected or most recently used user memory.

Unlocking the cuff and removing the device

After the measurement has been taken successfully, you can take off the device again. To do so, proceed as 

follows:

• P

• Y

Transfer of measurements via 

Bluetooth

®

In addition to displaying and saving your measurements locally on the device, you have the option of trans

-

ferring your measurements to your smartphone using 

Bluetooth

®

 low energy technology.

For this you need the “beurer HealthManager Pro” or “beurer HealthManager” app. These apps are available 

for free in the Apple App Store and from Google Play.

System requirements:
– iOS ≥  12.0, Android

TM

 ≥ 8.0

– 

Bluetooth

®

 ≥ 4.0

List of compatible devices:   

To transfer the measured values, proceed as follows:

Step 1: 

Activate the 

Bluetooth

 

®

 function on your device as described in section “Setting 

the hour format, date, time and 

Bluetooth

 

®

 

”.

Step 2:   

“beurer HealthManager Pro” or “beurer HealthManager” app

In the „beurer HealthManager Pro“ or „beurer HealthManager“ app, add the BM 81 

under Settings / Devices and follow the instructions.

Содержание BM 81 easyLock

Страница 1: ...EN Upper arm blood pressure monitor Instructions for use BM 81 easyLock 18 Model MD6000 ...

Страница 2: ...ates a hazardous situation which if not avoided could result in death or serious injury CAUTION indicates a hazardous situation which if not avoided may result in minor or moderate injury Contents Read these instructions for use carefully and keep them for later use be sure to make them accessible to other users and observe the information they contain Dear customer Thank you for choosing a produc...

Страница 3: ...atteries containing hazardous substances with household waste 21 PAP Packaging disposal Dispose of packaging in an environmentally friendly manner Manufacturer Authorized representative in the European Community Distributor Temperature limit The temperature limit values to which the medical device can safely be exposed are indicated Humidity limit Indicates the humidity range to which the medical ...

Страница 4: ...lood pressure monitor on newborns children or pets People with restricted physical sensory or mental skills should be supervised by a person responsible for their safety and receive instructions from this person on how to use the device If you have any of the following conditions it is essential you consult your doctor before using the device cardiac arrhythmia circulatory problems diabetes pregna...

Страница 5: ...tored close to the maximum or minimum storage and transport temperatures and is placed in an environment with a D o not drop the device D o not use the device in the vicinity of strong electromagnetic fields and keep it away from radio systems or mobile telephones W The temperature of cuff may reached to 42 1 C during normal operation at 40 0 C environment e recommend that the batteries be removed...

Страница 6: ...those specified or provided by the manufacturer of this device can lead to an increase in electromagnetic emissions or a decrease in the device s electromagnetic immunity this can result in faulty operation Failure to comply with the above can impair the performance of the device 5 Device description Blood pressure monitor and cuff M 3 7 2 1 8 6 5 4 1 Display 2 Upper arm cuff 3 easyLock locking bu...

Страница 7: ...e time and Bluetooth settings It is essential to set the date and time Otherwise you will not be able to save your measured values correctly with a date and time and access them again later Before initial us e and after each tim e you replace the ba ttery When inserting batteries into the device you will be taken to the relevant menu automatically If the batteri es have already been inserted Press...

Страница 8: ...order to generate as informative a profile of the progression of your blood pressure as possible and ensure that the measured values can be compared you should measure your blood pressure regularly and always at the same times of day It is recommended that you measure your blood pressure twice a day once in the morning after getting up and once in the evening Y fore avoid taking measurements durin...

Страница 9: ...it clicks into place to activate the lock ing mechanism A Lock locking button T T Cuff position control The device has a cuff position control function Using this feature the device automatically checks the correct positioning of the cuff on your upper arm during measurement If the cuff is correctly positioned the symbol will appear on the display during inflation It will remain there until the me...

Страница 10: ...and confirm your selection with the Start stop button Performing the blood pressure measurement Measurement To start the blood pressure monitor press the Start stop button A ll display elements are briefly displayed The blood pressure monitor will begin the measurement automatically after approx 3 seconds T the pulse symbol is displayed You can cancel the measurement at any time by pressing the St...

Страница 11: ...user memory Unlocking the cuff and removing the device After the measurement has been taken successfully you can take off the device again To do so proceed as follows P Y Transfer of measurements via Bluetooth In addition to displaying and saving your measurements locally on the device you have the option of trans ferring your measurements to your smartphone using Bluetooth low energy technology F...

Страница 12: ...s and blood is pumped into the blood vessels The lowest is diastolic blood pressure which is when the heart muscle has completely stretched back out and the heart fills with blood F information about the actual blood pressure Reliable assessment is only possible when you perform the measurement regularly under comparable conditions Irregular Heartbeat This device can identify any Irregular Heartbe...

Страница 13: ...y serve as a general guideline as the individual blood Furthermore it must be noted that measurements taken yourself while at home are generally lower than those that are taken by the doctor For this reason it is important that you regularly consult your doctor for advice Only they are able to give you your personal target values for controlled blood pressure in particular if you receive medicinal...

Страница 14: ... this user memory is displayed Press and hold both setting buttons and CL 00 appears on the display All the values in the current user memory are deleted Das Gerät schaltet sich im Anschluss automatisch ab 8 Cleaning and maintenance Clean the device and cuff carefully using a slightly damp cloth only Do not use any cleaning agents or solvents Under no circumstances hold the device and cuff under w...

Страница 15: ...t empty Insert new batteries into the device ER7 Unable to transfer the data via Bluetooth Please observe the information in chapter Transfer of measurements via Bluetooth 11 Disposal Repairing and disposing of the device Do not repair or adjust the device yourself Proper operation can no longer be guaranteed in this case Do not open the device Failure to comply will invalidate the warranty Repair...

Страница 16: ...atteries with cuff Cuff size 24 to 40 cm Permissible operating conditions 5 C to 40 C 15 90 relative air humidity non condensing Permissible storage and transport conditions 25 C to 70 C up to 90 non condensing Atmospheric pressure 700 hPa to 1060 hPa Power supply 4 x 1 5 V AAA batteries Battery life For approx 120 measurements depending on the blood pressure level and or pump pressure Classificat...

Страница 17: ...his product subject to the requirements below and to the extent described as follows from the sales agreement with the buyer The warranty shall apply without prejudice to any mandatory statutory provisions on liability Beurer guarantees the perfect functionality and completeness of this product The worldwide warranty period is 5 years commencing from the purchase of the new unused product from the...

Страница 18: ...isions of the instructions for use as well as products that have been opened repaired or modified by the buyer or by a service centre not authorised by Beurer damage that arises during transport between manufacturer and customer or between service centre and customer products purchased as seconds or as used goods consequential damage arising from a fault in this product however in this case claims...

Страница 19: ...f the Sphygmomanometer MD6000 should assure that it is used in such an environment RF emissions CISPR11 RF emissions CISPR11 Harmonic emissions IEC 61000 3 2 Voltage fluctuations flicker emissions IEC 61000 3 3 Group 1 Class B Class A Compliance Emissions test Compliance Guidance and manufacturer s declaration electromagnetic emissions ...

Страница 20: ...UT is the a c mians voltage prior to application of the test level 30 A m 50Hz 60Hz 30 A m 50Hz 60Hz Radiated RF IEC 61000 4 3 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz 10 V m 80 MHz 2 7 GHz 80 AM at 1 kHz Conduced RF IEC 61000 4 6 150KHz to 80MHz 3Vrms 6Vrms in ISM and amateur radio bands 80 Am at 1kHz 150KHz to 80MHz 3Vrms 6Vrms in ISM and amateur radio bands 80 Am at 1kHz Voltage dips short interrup...

Страница 21: ... 0 3 28 Pulse modulation 217 Hz 710 745 780 810 870 930 704 787 LTE Band 13 17 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 0 2 0 3 9 Pulse modulation 217 Hz Pulse modulation 18 Hz 0 2 0 3 9 5240 5500 5785 5100 5800 WLAN 802 11 a n Pulse modulation 217 Hz 2 0 3 28 1720 1845 1970 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation 217 Hz Radiated RF IEC6...

Страница 22: ...maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 10 V m 0 01 0 1 1 10 100 0 200 0 632 2 000 6 33 20 0 0 060 0 190 0 600 1 90 6 00 3 Vrms For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation app...

Страница 23: ...utlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Changes or modifications to the product not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment IC Regulations This device contains licence exempt transmitter s receiver s that comply with Innovat...

Страница 24: ...ons equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the Digital Automatic Blood Pressure Monitor model MD6000 including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result If any a list of all cables and maximum lengths of cables if applicable transduce...

Страница 25: ...d of the usage Please follow Local Ordinances or Regulations for disposal MADE IN CHINA version number 1 0 software version 1 0 0123 Grandway Technology Shenzhen Limited No 5 the Second Industrial Zone Zhukeng Community Longtian Street Pingshan District Shenzhen 518118 The People s Republic of China Shanghai International Holding Corp GmbH Europe Eiffestrasse 80 20537 Hamburg Germany ...

Отзывы: