Chapter 4. Software Functions
24
MicroPolar (++) LB 566
4.2.12 Massflow
The MASSFLOW menu appears only if two prerequisites have
been fulfilled:
1. The density unit g/cm^3 is selected for the concentration.
2. Current input 2 is enabled and m/s has been selected as
unit for current input 2.
For this case, the mass flow (throughput) in tons per hour
can be displayed and output via the current output, based on
the density reading with indication of the pipe cross-section
or the internal pipe diameter.
Massflow calculation:
Here the calculation is enabled or disabled.
Cross-sectional area:
Enter the cross-sectional area of the pipe.
Pipe inside diameter:
Enter the internal pipe diameter.
Comment: Only one entry is required: either the cross-
sectional area or the internal pipe diameter, the other pa-
rameter will be calculated automatically.
IMPORTANT
The massflow calculation is performed only at varying prod-
uct speed which must be entered via current input 2.
1 | - | Mass Flow | 07.05–13:25
Massfl. calculation
Enabled
Cross-sectional area
0.79cm^2
Pipe inside diameter
1.0cm
⌂◄
▲
▼
►
Содержание MicroPolar LB 566
Страница 1: ...Concentration Meters MicroPolar LB 566 User s Guide Hardware Manual 41986BA2 Rev Nr 06 E 09 2017 ...
Страница 2: ......
Страница 3: ......
Страница 16: ......
Страница 34: ......
Страница 50: ......
Страница 63: ...Chapter 7 Certificates MicroPolar LB 566 63 Certificates Chapter 7 7 1 EC Declaration of Conformity ...
Страница 64: ...Chapter 7 Certificates 64 MicroPolar LB 566 ...
Страница 65: ...Chapter 7 Certificates MicroPolar LB 566 65 7 2 Frequency License ...
Страница 66: ...Chapter 7 Certificates 66 MicroPolar LB 566 ...
Страница 67: ...Chapter 7 Certificates MicroPolar LB 566 67 ...
Страница 68: ...Chapter 7 Certificates 68 MicroPolar LB 566 ...
Страница 69: ...Chapter 7 Certificates MicroPolar LB 566 69 ...
Страница 70: ...Chapter 7 Certificates 70 MicroPolar LB 566 ...
Страница 71: ...Chapter 7 Certificates MicroPolar LB 566 71 ...
Страница 72: ...Chapter 7 Certificates 72 MicroPolar LB 566 ...
Страница 74: ...Chapter 8 Technical Drawings 74 MicroPolar LB 566 ...
Страница 77: ...Chapter 8 Technical Drawings MicroPolar LB 566 77 8 3 2 Type LB 5660 112 00X FlowCell DN 50 G BS M ...
Страница 79: ...Chapter 8 Technical Drawings MicroPolar LB 566 79 8 4 2 Type LB 5650 02 ...
Страница 80: ...Chapter 8 Technical Drawings 80 MicroPolar LB 566 8 4 3 Type LB 5650 03 ...
Страница 81: ...Chapter 8 Technical Drawings MicroPolar LB 566 81 8 4 4 Type LB 5650 04 ...
Страница 82: ...Chapter 8 Technical Drawings 82 MicroPolar LB 566 8 4 5 Type LB 5650 05 ...
Страница 83: ...Chapter 8 Technical Drawings MicroPolar LB 566 83 8 4 6 Type LB 5650 09 ...
Страница 84: ...Chapter 8 Technical Drawings 84 MicroPolar LB 566 8 4 7 Installation Situation in Pipelines ...
Страница 86: ...Chapter 8 Technical Drawings 86 MicroPolar LB 566 8 5 2 Type LB 5651 02 ...
Страница 87: ...Chapter 8 Technical Drawings MicroPolar LB 566 87 8 5 3 Type LB 5651 03 ...
Страница 88: ...Chapter 8 Technical Drawings 88 MicroPolar LB 566 8 5 4 Installation Situation in Pipelines ...
Страница 91: ...Notes MicroPolar LB 566 91 Notes ...
Страница 92: ...Notes 92 MicroPolar LB 566 Notes ...
Страница 93: ...Concentration Meters MicroPolar LB 566 User s Guide Software Manual 41986BA2 Rev Nr 06 E 09 2017 ...
Страница 94: ......
Страница 98: ......
Страница 100: ......
Страница 102: ......
Страница 104: ......
Страница 132: ......
Страница 146: ......
Страница 164: ......
Страница 168: ......
Страница 176: ......