Chapter 10. Error Lists and Device States
MicroPolar (++) LB 566
73
Chapter 10.
Error Lists and Device States
The LEDs indicate the device status. Once the errors have been
corrected, the measurement returns to the state before the er-
ror occurred. An acknowledgment is not required.
10.1 Hardware and Error Prompts
Code Error
Possible cause
14
Battery voltage
Battery power is low, replace im-
mediately,
see Hardware Manual, chapter 5.4
20
Caution!
Ambient tem-
perature too
high!
Check operating temperature of
the evaluation unit, permissible
range:
-20 to 60° C
21
HF temperature
out of range
Check operating temperature of
the evaluation unit, permissible
range:
-20 to 60° C
32
Parameter
memory
faulty
Compatibility check after software
download: A general reset must
be carried out.
39
HF hardware
failure
Faulty cable connection between
the motherboard and HF unit.
Check connector on the mother-
board.
Caution! First, disconnect the
evaluation unit from the power
supply!
For all other error messages, please contact the Berthold
Technologies Service.
10.2 Input Error
Error
Probable Cause
Value too large Input value is too large
Value too
small
Input value is too small
Table is empty
Sampling has been selected without
previous sample measurement
Chart data
faulty
The measuring system has determined
faulty chart data during calibration.
No chart data
available
The calculated chart data have been deleted
or calibration has not been completed.
Sampling full
You have tried to measure more than 30
samples.
Содержание MicroPolar LB 566
Страница 1: ...Concentration Meters MicroPolar LB 566 User s Guide Hardware Manual 41986BA2 Rev Nr 06 E 09 2017 ...
Страница 2: ......
Страница 3: ......
Страница 16: ......
Страница 34: ......
Страница 50: ......
Страница 63: ...Chapter 7 Certificates MicroPolar LB 566 63 Certificates Chapter 7 7 1 EC Declaration of Conformity ...
Страница 64: ...Chapter 7 Certificates 64 MicroPolar LB 566 ...
Страница 65: ...Chapter 7 Certificates MicroPolar LB 566 65 7 2 Frequency License ...
Страница 66: ...Chapter 7 Certificates 66 MicroPolar LB 566 ...
Страница 67: ...Chapter 7 Certificates MicroPolar LB 566 67 ...
Страница 68: ...Chapter 7 Certificates 68 MicroPolar LB 566 ...
Страница 69: ...Chapter 7 Certificates MicroPolar LB 566 69 ...
Страница 70: ...Chapter 7 Certificates 70 MicroPolar LB 566 ...
Страница 71: ...Chapter 7 Certificates MicroPolar LB 566 71 ...
Страница 72: ...Chapter 7 Certificates 72 MicroPolar LB 566 ...
Страница 74: ...Chapter 8 Technical Drawings 74 MicroPolar LB 566 ...
Страница 77: ...Chapter 8 Technical Drawings MicroPolar LB 566 77 8 3 2 Type LB 5660 112 00X FlowCell DN 50 G BS M ...
Страница 79: ...Chapter 8 Technical Drawings MicroPolar LB 566 79 8 4 2 Type LB 5650 02 ...
Страница 80: ...Chapter 8 Technical Drawings 80 MicroPolar LB 566 8 4 3 Type LB 5650 03 ...
Страница 81: ...Chapter 8 Technical Drawings MicroPolar LB 566 81 8 4 4 Type LB 5650 04 ...
Страница 82: ...Chapter 8 Technical Drawings 82 MicroPolar LB 566 8 4 5 Type LB 5650 05 ...
Страница 83: ...Chapter 8 Technical Drawings MicroPolar LB 566 83 8 4 6 Type LB 5650 09 ...
Страница 84: ...Chapter 8 Technical Drawings 84 MicroPolar LB 566 8 4 7 Installation Situation in Pipelines ...
Страница 86: ...Chapter 8 Technical Drawings 86 MicroPolar LB 566 8 5 2 Type LB 5651 02 ...
Страница 87: ...Chapter 8 Technical Drawings MicroPolar LB 566 87 8 5 3 Type LB 5651 03 ...
Страница 88: ...Chapter 8 Technical Drawings 88 MicroPolar LB 566 8 5 4 Installation Situation in Pipelines ...
Страница 91: ...Notes MicroPolar LB 566 91 Notes ...
Страница 92: ...Notes 92 MicroPolar LB 566 Notes ...
Страница 93: ...Concentration Meters MicroPolar LB 566 User s Guide Software Manual 41986BA2 Rev Nr 06 E 09 2017 ...
Страница 94: ......
Страница 98: ......
Страница 100: ......
Страница 102: ......
Страница 104: ......
Страница 132: ......
Страница 146: ......
Страница 164: ......
Страница 168: ......
Страница 176: ......