Uni-Probe LB 490
BERTHOLD TECHNOLOGIES GmbH & Co. KG
1 – 77
Volume 1
9 Functional Safety
1
©
exida.com
GmbH
berthold 0408-10 r003 v1r3.doc, Apr. 12, 2007
Rainer Faller
Page 9 of 18
3.2 Measuring
principle
Figure 2 gives an overview of the measuring principle of the Level Transmitter LB490 Uni-Probe.
Figure 2: Measuring principle of the Level Transmitter LB490 Uni-Probe
Explanation of terms:
Strahlenquelle:
Capsulated and shielded Gamma radiation source mounted outside the vessel
at the level to be monitored
Messsystem:
Radiation detector mounted opposite to the source
Ausgang:
4..20mA NAMUR NE43 compliant output
The Level Switch system is based on the physical law of the
attenuation
of
Gamma
radiation
as
it passes through medium. Product in between the source and
the
corresponding
detector
decreases the detected radiation by a related extent. This effect
corresponds
to
the
relative
product's
presence
and
therewith
it
signalizes
the level of the product in the container (vessel,
pipe, etc.).
As
it is a contact-less measurement with external mounting without modification of the existing
vessel, the measurement is independent of:
x
High temperature with water-cooling
x
High pressure or vacuum
x
Volatile & biohazard material
x
Corrosive material
x
Agitators, baffles, coils etc.
x
Build up on vessel walls
x
Physical and chemical properties of the product and the process
Berthold
Technologies main detector properties are designed according to the patented method
of
automatic
drift stabilization for radiometric applications, this method uses the energy loss of
natural cosmic radiation for readjustment of
the
amplification
for
measurement
result
corrections. (See also Patent scripture DE 41 14030 C1)
©
exida.com
GmbH
berthold 0408-10 r003 v1r3.doc, Apr. 12, 2007
Rainer Faller
Page 10 of 18
4
Failure Modes, Effects, and Diagnostics Analy
sis
The
Failure
Modes,
Effects, and Diagnostic Analysis was done together with Berthold
Technology
and
is
documented in [R1] to [R3]. When the effect of a certain failure mode could
not
be
analyzed
theoretically, the failure modes were introduced on component level and the
effects of these failure modes were examined on system level (see [D7] – Fault insertion
tests).
This resulted in failures that can be classified according to the following failure categories.
4.1
Description of the failure categories
In order to judge the failure behavior of the Level Transmitter LB490 Uni-Probe, the
following
definitions for the failure of the product were considered.
Fail-Safe
State
The fail-safe state is defined as the current contact reaching the
NAMUR NE43 Alarm ranges, i.e., I < 3,6mA or I > 21mA.
Fail
Safe
A safe failure (S) is defined as a failure that causes the transmitter
to
go
to
the
defined
fail-safe state without a demand from the
process. Safe failures are divided into safe detected (SD)
and
safe
undetected (SU) failures.
Fail
Dangerous
A
dangerous
failure
(D) is defined as a failure that does not
correctly
respond
to
a
demand
from the process outside a band of
more than 5% full span at ambient temperature.
Fail
Dangerous
Undetected
Failure that is dangerous and that is not
being
diagnosed
by
internal diagnostics.
Fail
Dangerous
Detected
Failure that is dangerous but is detected by
internal
diagnostics
or
a
connected
logic solver (These failures may be converted to the
selected fail-safe state).
Annunciation
Failure,
e.g.
in a diagnostic circuit, that does not directly impact
safety but impacts the ability to detect a future fault. Annunciation
failures are divided into annunciation detected (AD)
and
annunciation undetected (AU) failures. For the calculation of
the
Safe Failure Fraction (SFF), they are treated like dangerous
failures. This is a worst-case interpretation.
No
Effect
failures
Failure
of
a
component
that is part of the safety function but has
no
effect
on the safety function within a band of not more than 5%
full span. For the calculation of the Safe
Failure
Fraction
(SFF),
it
is treated like a safe undetected failure.
The
failure categories listed above expand on the categories listed in IEC 61508 which are only
safe
and
dangerous,
both
detected and undetected. The reason for this is that not all failure
modes
have
effects
that can be accurately classified according to the failure categories listed in
IEC 61508.
“No Effect” and “Annunciation” failures are provided for those who wish to do reliability
modeling
more
detailed
than
required by IEC 61508. In IEC 61508:2000, “No Effect” failures are defined
as safe undetected failures even though they will not cause the safety function to go to a safe
state. Therefore they need to be considered in the Safe Failure Fraction calculation.
Содержание Uni-Probe LB 490
Страница 2: ......
Страница 17: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 1 Safety Manuals...
Страница 18: ...38477BA2B 1 18 25 5 09 Volume 1...
Страница 22: ...38477BA2B 1 22 25 5 09 1 About this User s Manual Volume1...
Страница 28: ...38477BA2B 1 28 25 5 09 4 Transport and Assembly Volume1...
Страница 44: ...38477BA2B 1 44 25 5 09 6 Radiation Protection Volume1...
Страница 56: ...38477BA2B 1 56 25 5 09 7 Source Replacement Volume1...
Страница 82: ...38477BA2B 1 82 25 5 09 9 Functional Safety Volume1...
Страница 90: ...38477BA2B 1 90 25 5 09 11 Certificates Volume1...
Страница 91: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 91 Volume 1 11 Certificates 1...
Страница 92: ...38477BA2B 1 92 25 5 09 11 Certificates Volume1...
Страница 93: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 93 Volume 1 11 Certificates 1...
Страница 94: ...38477BA2B 1 94 25 5 09 11 Certificates Volume1...
Страница 95: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 95 Volume 1 11 Certificates 1...
Страница 96: ...38477BA2B 1 96 25 5 09 11 Certificates Volume1...
Страница 97: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 97 Volume 1 11 Certificates 1...
Страница 98: ...38477BA2B 1 98 25 5 09 11 Certificates Volume1...
Страница 99: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 99 Volume 1 11 Certificates 1...
Страница 100: ...38477BA2B 1 100 25 5 09 11 Certificates Volume1...
Страница 101: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 101 Volume 1 11 Certificates 1...
Страница 102: ...38477BA2B 1 102 25 5 09 11 Certificates Volume1...
Страница 103: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 103 Volume 1 11 Certificates 1...
Страница 104: ...38477BA2B 1 104 25 5 09 11 Certificates Volume1...
Страница 105: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 105 Volume 1 11 Certificates 1...
Страница 106: ...38477BA2B 1 106 25 5 09 11 Certificates Volume1...
Страница 107: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 107 Volume 1 11 Certificates 1...
Страница 108: ...38477BA2B 1 108 25 5 09 11 Certificates Volume1 11 2 FM Certificate...
Страница 109: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 109 Volume 1 11 Certificates 1...
Страница 112: ...38477BA2B 1 112 25 5 09 11 Certificates Volume1 11 4 EG Declaration of Conformity...
Страница 115: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 1 115 Volume 1 1 Notes...
Страница 117: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 2 Uni Probe Installation...
Страница 118: ...38477BA2B 2 118 25 5 09 Volume 2...
Страница 138: ...38477BA2B 2 138 25 5 09 1 System Description Volume2...
Страница 254: ...38477BA2B 2 254 25 5 09 6 Technical Drawings Volume2 6 2 Rod Detector...
Страница 286: ...38477BA2B 2 286 25 5 09 6 Technical Drawings Volume2 6 10 Shielding for Rod Source on Dip Tube...
Страница 288: ...38477BA2B 2 288 25 5 09 6 Technical Drawings Volume2...
Страница 292: ...38477BA2B 2 292 25 5 09 7 Cooling Water Curves Volume2...
Страница 293: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 2 293 Volume 2 2 Notes...
Страница 295: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 3 HART Communicator User Interface...
Страница 296: ...38477BA2B 3 296 25 5 09 Volume 3...
Страница 340: ...38477BA2B 3 340 25 5 09 3 Getting Started via the HART Communicator Volume3...
Страница 403: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 3 403 Volume 3 3 Notes...
Страница 405: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 4 PACTware User Interface FDT DTM...
Страница 406: ...38477BA2B 4 406 25 5 09 Volume 4...
Страница 408: ...38477BA2B 4 408 25 5 09 1 PC Connection to the Uni Probe Volume4...
Страница 416: ...38477BA2B 4 416 25 5 09 2 Installing and Working with DTM Volume4...
Страница 417: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 4 417 Volume 4 4 Notes...
Страница 419: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 5 SIMATIC PDM User Interface HART...
Страница 422: ...38477BA2B 5 422 25 5 09 1 Hardware Configuration Volume5...
Страница 490: ...38477BA2B 5 490 25 5 09 4 Menu Overview Volume5...
Страница 512: ...38477BA2B 5 512 25 5 09 5 Calibration with SIMATIC PDM Volume5...
Страница 526: ...38477BA2B 5 526 25 5 09 7 Working with SIMATIC PDM Volume5...
Страница 542: ...38477BA2B 5 542 25 5 09 8 Explanations Volume5...
Страница 550: ...38477BA2B 5 550 25 5 09 9 Error Handling Volume5...
Страница 551: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 5 551 Volume 5 5 Notes...
Страница 553: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 6 SIMATIC PDM User Interface Profibus PA...
Страница 600: ...38477BA2B 6 600 25 5 09 3 Device specific Menus Volume6...
Страница 624: ...38477BA2B 6 624 25 5 09 4 Calibration with SIMATIC PDM Volume6...
Страница 648: ...38477BA2B 6 648 25 5 09 6 Explanations Volume6...
Страница 654: ...38477BA2B 6 654 25 5 09 7 Error Handling Volume6...
Страница 655: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 6 655 Volume 6 6 Notes...
Страница 657: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 7 FOUNDATION Fieldbus User Interface...
Страница 664: ...38477BA2B 7 664 25 5 09 2 Installation Program Start Volume7...
Страница 672: ...38477BA2B 7 672 25 5 09 3 Parameter Overview Volume7...
Страница 688: ...38477BA2B 7 688 25 5 09 4 Calibration with FOUNDATION Fieldbus Volume7...
Страница 700: ...38477BA2B 7 700 25 5 09 5 Functional Processes Volume7...
Страница 720: ...38477BA2B 7 720 25 5 09 7 Error Handling Volume7...
Страница 721: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 7 721 Volume 7 7 Notes...
Страница 728: ...Index Volume 1 7 38477BA2B I 728 25 5 09...
Страница 729: ...Uni Probe LB 490 BERTHOLD TECHNOLOGIES GmbH Co KG 729 Volume 1 7 Notes...